Lines of code count not available
fixed in latest code
implemented
fixed
I think this would be a TD s/w fault (alarm already define). Just need a new TD s/w fault ID in AlarmMgmtSWFaults.h.
This structure doesn't seem necessary - it's just a middle man. Why not just pass the #defines directly into the quadratic macro?
Fixed
removed from here
We need new alarm . That should be included in SRS right ? or shall i use SET_ALARM_WITH_2_U32_DATA( ALARM_ID_TD_SOFTWARE_FAULT) ?
Add blank line after declarations.
Alarm should be handled inside of setAirPumpState().
I think alarm should be handled inside the setAirPumpState() function.
Should we have alarm here?
duty misspelled.
Can we remove inclusion of fpga here? It shouldn't be needed.
pushed new code
Got a suggestion or an issue? Discuss it on the Atlassian Community or visit Atlassian Support for other options.