To resolve the build failure with DD, haven't added all the features, instead added TODO comment

To resolve the build failure with DD, haven't added all the features, instead added TODO comment

Idea is to have this in institutional records we implement it

Idea is to have this in institutional records we implement it

a time out for decaying D66 pressure to 3 PSI after D65 closed

a time out for decaying D66 pressure to 3 PSI after D65 closed

If this is const, how will user add conversions (after 3rd value)?

If this is const, how will user add conversions (after 3rd value)?

Remove extra blank line.

Remove extra blank line.

remove space after <= and add space end of function call.

remove space after <= and add space end of function call.

Is this time out after D80 opening or D65 closing state? if it is D65 closed, you need to start the timer after D65 closing.

Is this time out after D80 opening or D65 closing state? if it is D65 closed, you need to start the timer after D65 closing.

if we are not checking level, this can be removed and function name also can be updated accordingly.

if we are not checking level, this can be removed and function name also can be updated accordingly.

Is this a TODO? If so, add TODO to the comment and maybe be a little more descriptive of what needs to be done.

Is this a TODO? If so, add TODO to the comment and maybe be a little more descriptive of what needs to be done.

can we get 'runPIController' inside deadband control check?

can we get 'runPIController' inside deadband control check?

Can we add sodium and bicarb overrides now, so that we don't want to come back again for this change? Also, please get the dialin updated to reflect the new parameters additions in tdproxy and let ...

Can we add sodium and bicarb overrides now, so that we don't want to come back again for this change?
Also, please get the dialin updated to reflect the new parameters additions in tdproxy and let system team be in loop to update the required system scripts?

First two logics ANDed correct (first cycle and close period). the third one (BC switch only must be ANDed only with close period), not to include first cycle..

First two logics ANDed correct (first cycle and close period). the third one (BC switch only must be ANDed only with close period), not to include first cycle..

LEAHI-DD-FIRMWARE-LDT-4188_SW/FW integration to run a Treatment from UI
LEAHI-DD-FIRMWARE-LDT-4188_SW/FW integration to run a Treatment from UI
Can we check payload.index is equal to D6_LEVL (instead of above comparision)?

Can we check payload.index is equal to D6_LEVL (instead of above comparision)?

LEAHI-DD-FIRMWARE-LDT-3975_Water Treatment and Degassing - FW Iteration - 4/4:
LEAHI-DD-FIRMWARE-LDT-3975_Water Treatment and Degassing - FW Iteration - 4/4:
Remove space after 'Supply'.

Remove space after 'Supply'.

Align the value declaration indentation.

Align the value declaration indentation.

looks you are already checking this state in a line above.

looks you are already checking this state in a line above.

So this unit manages the sending of record messages, right? I don't see a unit managing read/write/erase jobs. Where is that queue?

So this unit manages the sending of record messages, right? I don't see a unit managing read/write/erase jobs. Where is that queue?

I think this enum should be in a different header file. I don't think this unit uses this enum.

I think this enum should be in a different header file. I don't think this unit uses this enum.

Slot size and offsets can probably be private and maybe better placed in a different header file? I don't see why a msg Q would need/use/be responsible for this kind of thing.

Slot size and offsets can probably be private and maybe better placed in a different header file? I don't see why a msg Q would need/use/be responsible for this kind of thing.

These addresses should be calculated in init function from sector start addresses (call driver get function to get these) and your offsets and stored in static variables within the .c file.

These addresses should be calculated in init function from sector start addresses (call driver get function to get these) and your offsets and stored in static variables within the .c file.

Do these need to be public? I think only the driver should have to know sector addresses and sizes. If other units want to know these things, they should call a function in this driver to get them.

Do these need to be public?
I think only the driver should have to know sector addresses and sizes.
If other units want to know these things, they should call a function in this driver to get them.

I think isSectorStartAddress is a better function name.

I think isSectorStartAddress is a better function name.

Be careful with bufferSize. Make sure you understand what it means (e.g. is it bytes or words). I believe the FAPI functions deal in 4-byte words and it caused some issues for Dara because he thoug...

Be careful with bufferSize. Make sure you understand what it means (e.g. is it bytes or words). I believe the FAPI functions deal in 4-byte words and it caused some issues for Dara because he thought this was bytes.

don't forget to add function headers.

don't forget to add function headers.

// **************** private functions **************

// **************** private functions **************