Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
LEAHI-TD-FIRMWARE-LDT-1903_Air Trap Level Control - TD - 04: DEV - Feature Implementation
LEAHI-TD-FIRMWARE-LDT-1903_Air Trap Level Control - TD - 04: DEV - Feature Implementation
LDT-1903 Air Trap H12 off during fill event

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch LDT-1903-air-trap-level-control-delay-adjust-after-fill-fix---td---0 into staging

LDT-1903 Air Trap reverting to original valve state

reverted back

reverted back

This code is removed and not required

This code is removed and not required

LDT-1903 Air Trap Final implementation

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
Remove spaces after if/else if.

Can we remove the empty lines after if / else if blocks in this function?

Can we remove the empty lines after if / else if blocks in this function?

Minor formatting and logic cleanup.

  1. … 1 more file in changeset.
You are calling setAirPumpState in if and else. Consider just calling once after the if/else.

You are calling setAirPumpState in if and else. Consider just calling once after the if/else.

LDT-3600: Skip air trap fill, lowerevents and venous bubble alarm when testing without tubing set.

  1. … 1 more file in changeset.
Add space between if and (

Add space between if and (

Fixed

Fixed

LEAHI-TD-FIRMWARE-LDT-3600_Request for TD Test Config for Testing without Tubing Set on TD
LEAHI-TD-FIRMWARE-LDT-3600_Request for TD Test Config for Testing without Tubing Set on TD
LDT-1903 Air Trap changing the first lowering event duty cycle

Comment to right no longer makes sense.

Comment to right no longer makes sense.

LDT-1907 IAr trap FW uses the stabilised H14 long filtered pressure right before the AP fill event

LDT-1903 Air Trap Addressed the review comments

LDT-1903 Air Trap -both H13 and H20 valves are closed at the end of the fill during the 1 sec delay before the AP turns on to lower the level

Should this event be TD_EVENT_AIR_TRAP_FILL?

Should this event be TD_EVENT_AIR_TRAP_FILL?

LDT-1903 Air Trap fixing the air pump turn on after 1 sec when raise event is completed