If you change this line to "if ( ++airPumpStallCounter >= AIR_PUMP_STALL_PERSISTENCE )", you can remove the if statement above.

If you change this line to "if ( ++airPumpStallCounter >= AIR_PUMP_STALL_PERSISTENCE )", you can remove the if statement above.

I was thinking that comment should go above the first if statement.

I was thinking that comment should go above the first if statement.

Updated

Updated

Define what 3 means (#define up top). If 3 means 3 x 50ms general task interval, define it as shown below: #define AIR_PUMP_STALL_PERSISTENCE ( 150 / TASK_GENERAL_INTERVAL ) ///< How long stall con...

Define what 3 means (#define up top). If 3 means 3 x 50ms general task interval, define it as shown below:
#define AIR_PUMP_STALL_PERSISTENCE ( 150 / TASK_GENERAL_INTERVAL ) ///< How long stall condition must persist before alarming.

OR use persistent alarm.

Added

Added

Add a comment explaining what we're looking for here (i.e. we are commanding air pump to run but zero measured pump speed indicates it is not running).

Add a comment explaining what we're looking for here (i.e. we are commanding air pump to run but zero measured pump speed indicates it is not running).

Added

Added

Should probably initialize your new stall counter variable.

Should probably initialize your new stall counter variable.

Sean Nash I need your help for writing this Alarm properties.

Sean Nash I need your help for writing this Alarm properties.

Removed

Removed

Updated

Updated

Still not packed.

Still not packed.

Remove extra blank line.

Remove extra blank line.

Sure

Sure

Can we remove the empty lines after if / else if blocks in this function?

Can we remove the empty lines after if / else if blocks in this function?

LEAHI-TD-FIRMWARE-LDT-3600_Request for TD Test Config for Testing without Tubing Set on TD
LEAHI-TD-FIRMWARE-LDT-3600_Request for TD Test Config for Testing without Tubing Set on TD
Removed

Removed

I don't think we need to call setPressureLimitsToOuterBounds() either. The signal function will indirectly cause that function to be called later.