Remove extra blank line.

Remove extra blank line.

Should we create a structure for these moving average filters and make an array of them based on enum of temp sensors? It looks kind of messy having separate handling for each sensor - can we conso...

Should we create a structure for these moving average filters and make an array of them based on enum of temp sensors?
It looks kind of messy having separate handling for each sensor - can we consolidate some of this code?

Should we maybe just have a constant array of sample counts based on enum of temp sensors?

Should we maybe just have a constant array of sample counts based on enum of temp sensors?

Should min and max be properties of the controller instead of global (assumes all controllers will want same min/max)? Controller properties already include a min and max field - can we use those?

Should min and max be properties of the controller instead of global (assumes all controllers will want same min/max)?
Controller properties already include a min and max field - can we use those?

Remove blank line.

Remove blank line.

Done.

Done.

Done.

Done.

mis aligned ?

mis aligned ?

mis aligned ?

mis aligned ?

need a brief description about this feature and how it is implemented/logic/math etc - also comment on the previous logic

need a brief description about this feature and how it is implemented/logic/math etc - also comment on the previous logic

its a reverse logic - is it OK to make positive instead of not defined ?

its a reverse logic - is it OK to make positive instead of not defined ?

Looks like we are doing a lot of the dialysate temp control in Gen Dialysate mode. Do we not do any heating in PreGen mode? Even if all dialysate heating is done in this mode, should we consider cr...

Looks like we are doing a lot of the dialysate temp control in Gen Dialysate mode. Do we not do any heating in PreGen mode?
Even if all dialysate heating is done in this mode, should we consider creating a service controller (like BC controller) to manage this complex temperature control?

Resolve merge conflicts.

Resolve merge conflicts.

Add blank line between declarations and rest of code.

Add blank line between declarations and rest of code.

Add comment to right listing which math functions we're using.

Add comment to right listing which math functions we're using.

LEAHI-DD-FIRMWARE-LDT-3352_DD - revisit D3 level control logic
LEAHI-DD-FIRMWARE-LDT-3352_DD - revisit D3 level control logic
Done.

Done.

Removed.

Removed.

Done.

Done.

Done.

Done.

Same comment - pend transition until BC cycle completes.

Same comment - pend transition until BC cycle completes.

Consider keeping the transition to spent chamber fill pending until balancing chamber cycle completes.

Consider keeping the transition to spent chamber fill pending until balancing chamber cycle completes.

Is this the right place to close the BC valves? Maybe it should be in transitionToBalChamberFill()? If we come back to this bypass state from delivery state, is it appropriate to close BC valves here?

Is this the right place to close the BC valves? Maybe it should be in transitionToBalChamberFill()?
If we come back to this bypass state from delivery state, is it appropriate to close BC valves here?

Align comment.

Align comment.

DD-LDT-1873_Dialysate Flow Rate DD
DD-LDT-1873_Dialysate Flow Rate DD
Done.

Done.