Is it an enum/index like Denali then?

Is it an enum/index like Denali then?

LDT-3676 LDT-3677 merge latest from staging

No we are not getting conv factor from UI on it I will have to confirm with Nick to see if we will get the bicarbConvFactor

No we are not getting conv factor from UI on it
I will have to confirm with Nick to see if we will get the bicarbConvFactor

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

LDT-1477 merge to staging

Align comments.

Align comments.

Can we just remove this?

Can we just remove this?

Can we just remove this?

Can we just remove this?

F suffix.

F suffix.

Is bicarbonate also a conversion factor?

Is bicarbonate also a conversion factor?

Can we just remove the old enum now?

Can we just remove the old enum now?

Not sure these min/max are right here either.

Not sure these min/max are right here either.

So we have all temp sensors getting averaged now, right? So should we have one get function that takes a sensor ID as param?

So we have all temp sensors getting averaged now, right? So should we have one get function that takes a sensor ID as param?

Remove extra blank lines.

Remove extra blank lines.

LEAHI-TD-FIRMWARE-LDT-473_Treatment Parameters (Create Rx) - TD - 04: DEV - Feature Implementation
LEAHI-TD-FIRMWARE-LDT-473_Treatment Parameters (Create Rx) - TD - 04: DEV - Feature Implementation
Done.

Done.

Done

Done

Done.

Done.

Same comment.

Same comment.

Having much larger P coeff is likely to cause a lot of oscillations and very slow to get to stable target.

Having much larger P coeff is likely to cause a lot of oscillations and very slow to get to stable target.

Input is d78TempAvgC

Input is d78TempAvgC

I don't think these min/max (for control output in output units i.e. PWM) are appropriate for limiting maximum error sum step size (error is in feedback/target units i.e. RPM).