Align comments.

Align comments.

Can we just remove this?

Can we just remove this?

Can we just remove this?

Can we just remove this?

F suffix.

F suffix.

Is bicarbonate also a conversion factor?

Is bicarbonate also a conversion factor?

Can we just remove the old enum now?

Can we just remove the old enum now?

LEAHI-TD-FIRMWARE-LDT-473_Treatment Parameters (Create Rx) - TD - 04: DEV - Feature Implementation
LEAHI-TD-FIRMWARE-LDT-473_Treatment Parameters (Create Rx) - TD - 04: DEV - Feature Implementation
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

LEAHI-TD-FIRMWARE-LDT-2749_Ability to Enable Air Bubble Alarm in Standby Mode
LEAHI-TD-FIRMWARE-LDT-2749_Ability to Enable Air Bubble Alarm in Standby Mode
LDT-2749 added the override function declaration for enabling venous bubble alarm in standby mode

Updated

Updated

Added

Added

If you change this line to "if ( ++airPumpStallCounter >= AIR_PUMP_STALL_PERSISTENCE )", you can remove the if statement above.

If you change this line to "if ( ++airPumpStallCounter >= AIR_PUMP_STALL_PERSISTENCE )", you can remove the if statement above.

I was thinking that comment should go above the first if statement.

I was thinking that comment should go above the first if statement.

Updated

Updated

Define what 3 means (#define up top). If 3 means 3 x 50ms general task interval, define it as shown below: #define AIR_PUMP_STALL_PERSISTENCE ( 150 / TASK_GENERAL_INTERVAL ) ///< How long stall con...

Define what 3 means (#define up top). If 3 means 3 x 50ms general task interval, define it as shown below:
#define AIR_PUMP_STALL_PERSISTENCE ( 150 / TASK_GENERAL_INTERVAL ) ///< How long stall condition must persist before alarming.

OR use persistent alarm.

Added

Added

Add a comment explaining what we're looking for here (i.e. we are commanding air pump to run but zero measured pump speed indicates it is not running).

Add a comment explaining what we're looking for here (i.e. we are commanding air pump to run but zero measured pump speed indicates it is not running).

Added

Added

Should probably initialize your new stall counter variable.

Should probably initialize your new stall counter variable.

Sean Nash I need your help for writing this Alarm properties.

Sean Nash I need your help for writing this Alarm properties.