Why do some functions take sensor as a U32 and some as a CONDUCTIVITY_SENSORS_T? The latter seems appropriate.

Why do some functions take sensor as a U32 and some as a CONDUCTIVITY_SENSORS_T? The latter seems appropriate.

Remove blank line.

Remove blank line.

Remove blank line.

Remove blank line.

Add blank line before return.

Add blank line before return.

Add blank line before return.

Add blank line before return.

Remove blank line.

Remove blank line.

Not sure these min/max are right here either.

Not sure these min/max are right here either.

So we have all temp sensors getting averaged now, right? So should we have one get function that takes a sensor ID as param?

So we have all temp sensors getting averaged now, right? So should we have one get function that takes a sensor ID as param?

Remove extra blank lines.

Remove extra blank lines.

Done.

Done.

Done

Done

Done.

Done.

Same comment.

Same comment.

Having much larger P coeff is likely to cause a lot of oscillations and very slow to get to stable target.

Having much larger P coeff is likely to cause a lot of oscillations and very slow to get to stable target.

Input is d78TempAvgC

Input is d78TempAvgC

I don't think these min/max (for control output in output units i.e. PWM) are appropriate for limiting maximum error sum step size (error is in feedback/target units i.e. RPM).

I don't think these min/max (for control output in output units i.e. PWM) are appropriate for limiting maximum error sum step size (error is in feedback/target units i.e. RPM).

Done.

Done.

Just following the current style of code implementation to keep it uniform. These defines are temporary and will be removed.

Just following the current style of code implementation to keep it uniform. These defines are temporary and will be removed.

Done.

Done.

Done.

Done.

Done.

Done.

LEAHI-DD-FIRMWARE-LDT-2008_Dialysate Composition - DD - 04: DEV - Feature Implementation
LEAHI-DD-FIRMWARE-LDT-2008_Dialysate Composition - DD - 04: DEV - Feature Implementation
LEAHI-DD-FIRMWARE-LDT-2004_Dialysate Composition - DD
LEAHI-DD-FIRMWARE-LDT-2004_Dialysate Composition - DD
LEAHI-DD-FIRMWARE-LDT-2004_Dialysate Composition - DD
LEAHI-DD-FIRMWARE-LDT-2004_Dialysate Composition - DD
LEAHI-DD-FIRMWARE-LDT-2004_Dialysate Composition - DD
LEAHI-DD-FIRMWARE-LDT-2004_Dialysate Composition - DD