LEAHI-DD-FIRMWARE-LDT-4873_(Request) Add config to disable conductive level sensor fault
LEAHI-DD-FIRMWARE-LDT-4873_(Request) Add config to disable conductive level sensor fault
LDT-3735: Updated minimum dialysate pump flow rate to 50

LEAHI-DD-FIRMWARE-LDT-3902_Allow DD balancing chamber switch only command to go down to 50 mL/min
LEAHI-DD-FIRMWARE-LDT-3902_Allow DD balancing chamber switch only command to go down to 50 mL/min
Same as above.

Same as above.

Add an event here, when time out happens. Event data should capture fresh side and spent side fill completion status.

Add an event here, when time out happens. Event data should capture fresh side and spent side fill completion status.

You may need to comment/ remove this condition. the pressure stabilized flag should set only when pressure is met (not on chamber fill request)

You may need to comment/ remove this condition. the pressure stabilized flag should set only when pressure is met (not on chamber fill request)

LEAHI-DD-FIRMWARE-LDT-4612_Dialysate Composition (Dry Bicarbonate) - FW Iteration - 4/1:
LEAHI-DD-FIRMWARE-LDT-4612_Dialysate Composition (Dry Bicarbonate) - FW Iteration - 4/1:
it shouldn't be the default value

it shouldn't be the default value

Bamboo Commit: Updated DDCommon.h with build versions from Bamboo.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    /firmware/App/Controllers/MixingControl.c
Merge branch LDT-4612-dialysate-composition-integ into staging

LDT-4612 depressurize sanity test updates

    • -11
    • +0
    /firmware/App/Controllers/DryBiCart.c
Merge branch LDT-4612-dialysate-composition-integ into staging

LDT-4612 mixing control updates

    • -0
    • +2
    /firmware/App/Controllers/MixingControl.c
LDT-4612 depressurize updates

Merge branch LDT-4612-dialysate-composition-integ into staging

LDT-4612 reseting the top level state machine when depressurize completed

LEAHI-DD-FIRMWARE-LDT-4734_Water Treatment and Degassing - FW Iteration - 5/4:
LEAHI-DD-FIRMWARE-LDT-4734_Water Treatment and Degassing - FW Iteration - 5/4:
LEAHI-DD-FIRMWARE-LDT-4733_(Request) Add variables to capture AC/DC heater electrical data
LEAHI-DD-FIRMWARE-LDT-4733_(Request) Add variables to capture AC/DC heater electrical data
LDT-4612 addressed review comments

    • -2
    • +13
    /firmware/App/Controllers/DryBiCart.c
LDT-4612 addressed review comments

    • -7
    • +61
    /firmware/App/Controllers/DryBiCart.c
LDT-4612 addressed review comments

    • -80
    • +72
    /firmware/App/Controllers/DryBiCart.c
    • -1
    • +1
    /firmware/App/Controllers/MixingControl.c
Space after if.

Space after if.

Can we bring depressurize as well one of the states like fill, supply and drain? And the same way when one of the operations initiated and other should not kick in.

Can we bring depressurize as well one of the states like fill, supply and drain? And the same way when one of the operations initiated and other should not kick in.

remove empty space after d66pressure variable. or match the indent to previous statement.

remove empty space after d66pressure variable. or match the indent to previous statement.

Add a TODO comment (remove later): this should be handled in IO/FP module.

Add a TODO comment (remove later): this should be handled in IO/FP module.