Controllers

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Can we be a little more specific about what conductivity we are returning in these functions? And I think we should include constants in the Inputs section of function headers.

Can we be a little more specific about what conductivity we are returning in these functions?
And I think we should include constants in the Inputs section of function headers.

LEAHI-DD-FIRMWARE-LDT-4614_Dialysate Composition (Dry Bicarbonate) - FW Implementation - 5/1:
LEAHI-DD-FIRMWARE-LDT-4614_Dialysate Composition (Dry Bicarbonate) - FW Implementation - 5/1:
updated

updated

updated

updated

updated

updated

moved.

moved.

updated

updated

updated

updated

Move below Messaging.h to keep list in alphabetical order.

Move below Messaging.h to keep list in alphabetical order.

balancingError is an input.

balancingError is an input.

Same here. Let's use only D29.

Same here. Let's use only D29.

Suppose to use D74, Instead of D17 (for Control only).

Suppose to use D74, Instead of D17 (for Control only).

Move these functions to top area where all public functions defined. let's keep the static function after the public function.

Move these functions to top area where all public functions defined. let's keep the static function after the public function.

Missing Comment.

Missing Comment.

Add previousHeaterState to input and output

Add previousHeaterState to input and output

add prevDryBicartState to input and output.

add prevDryBicartState to input and output.

Add prevBalChamberState to input and output

Add prevBalChamberState to input and output

Add isBalChamberSwitchingActive to input

Add isBalChamberSwitchingActive to input

yes

yes

yes

yes

yes

yes

Align it with declration.

Align it with declration.

Add comments.

Add comments.

LEAHI-DD-FIRMWARE-LDT-4719_Dialysate Composition (Mixing) - FW Implementation - 4/1:
LEAHI-DD-FIRMWARE-LDT-4719_Dialysate Composition (Mixing) - FW Implementation - 4/1:
LEAHI-DD-FIRMWARE-LDT-4873_(Request) Add config to disable conductive level sensor fault
LEAHI-DD-FIRMWARE-LDT-4873_(Request) Add config to disable conductive level sensor fault
LDT-3735: Updated minimum dialysate pump flow rate to 50