What alarm is this? It doesn't look like you've added a new alarm. This is one of the reserved (unused) alarms that UI s/w might use.

What alarm is this? It doesn't look like you've added a new alarm. This is one of the reserved (unused) alarms that UI s/w might use.

Do we really need these setup functions for each state? They all look very simple - just 1 or 2 lines of code.

Do we really need these setup functions for each state? They all look very simple - just 1 or 2 lines of code.

LEAHI-TD-FIRMWARE-LDT-4040_Rinseback - FW Implementation - 3/10:
LEAHI-TD-FIRMWARE-LDT-4040_Rinseback - FW Implementation - 3/10:
I believe the SyringePump controller will trigger this alarm at end of seek operation.

I believe the SyringePump controller will trigger this alarm at end of seek operation.

Syringe pump should already be stopped if preload is done. Do we need to call this here?

Syringe pump should already be stopped if preload is done. Do we need to call this here?

This is called from preload state. Not needed here I think.

This is called from preload state. Not needed here I think.

Don't put comments on function prototypes. Doxygen gets details of functions from the function header.

Don't put comments on function prototypes. Doxygen gets details of functions from the function header.

Why (void)?

Why (void)?

Where is this being called? Should be called from pre-treatment init I think.

Where is this being called? Should be called from pre-treatment init I think.

Confirmation signal sounds like a message handler from UI. Message handlers must accept a message as parameter and return BOOL. They must also be added to the message handler table in Messaging.c.

Confirmation signal sounds like a message handler from UI. Message handlers must accept a message as parameter and return BOOL. They must also be added to the message handler table in Messaging.c.

Why are we signaling syringe load confirmed here?

Why are we signaling syringe load confirmed here?

Do these states need to be public?

Do these states need to be public?

Move up one row to alphabetize.

Move up one row to alphabetize.

If this reserved2 is a U32, next register would be at 288 (not 285). Should we make reserved2 a U08?

If this reserved2 is a U32, next register would be at 288 (not 285). Should we make reserved2 a U08?

change accordingly

change accordingly

Rename to AlarmTableResposnePayload

Rename to AlarmTableResposnePayload

I would keep blank line between library includes and regular includes.

I would keep blank line between library includes and regular includes.

We can't have 2 alarms with ID 196. The enum only has the empty saline bag alarm. Looks like alarm 197 is duplicate of this 196 alarm you've added. I think you should just remove this new alarm 196...

We can't have 2 alarms with ID 196. The enum only has the empty saline bag alarm. Looks like alarm 197 is duplicate of this 196 alarm you've added.
I think you should just remove this new alarm 196 that you've added here.

Do we still need this?

Do we still need this?

test instead of handle

test instead of handle

Align =

Align =

Should modality be an enum type?

Should modality be an enum type?