Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
LDT-3215-measured-venous-pressure-decreases

  1. … 4 more files in changeset.
LEAHI-TD-FIRMWARE-LDT-3215_Measured venous pressure decreases to atmosphere if H12 air pump is dead headed
LEAHI-TD-FIRMWARE-LDT-3215_Measured venous pressure decreases to atmosphere if H12 air pump is dead headed
Updated

Updated

Added

Added

LDT-3215-measured-venous-pressure-decreases

If you change this line to "if ( ++airPumpStallCounter >= AIR_PUMP_STALL_PERSISTENCE )", you can remove the if statement above.

If you change this line to "if ( ++airPumpStallCounter >= AIR_PUMP_STALL_PERSISTENCE )", you can remove the if statement above.

I was thinking that comment should go above the first if statement.

I was thinking that comment should go above the first if statement.

Updated

Updated

LDT-3215-measured-venous-pressure-decreases

Define what 3 means (#define up top). If 3 means 3 x 50ms general task interval, define it as shown below: #define AIR_PUMP_STALL_PERSISTENCE ( 150 / TASK_GENERAL_INTERVAL ) ///< How long stall con...

Define what 3 means (#define up top). If 3 means 3 x 50ms general task interval, define it as shown below:
#define AIR_PUMP_STALL_PERSISTENCE ( 150 / TASK_GENERAL_INTERVAL ) ///< How long stall condition must persist before alarming.

OR use persistent alarm.

Added

Added

LDT-3215-measured-venous-pressure-decreases

Add a comment explaining what we're looking for here (i.e. we are commanding air pump to run but zero measured pump speed indicates it is not running).

Add a comment explaining what we're looking for here (i.e. we are commanding air pump to run but zero measured pump speed indicates it is not running).

Added

Added

LDT-3215-measured-venous-pressure-decreases

Should probably initialize your new stall counter variable.

Should probably initialize your new stall counter variable.

Sean Nash I need your help for writing this Alarm properties.

Sean Nash I need your help for writing this Alarm properties.

LDT-3215-measured-venous-pressure-decreases

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 8 more files in changeset.
fixed in latest code

fixed in latest code

implemented

implemented

fixed

fixed

LDT-1903 Air Trap adding duty cycle out of range software fault alarm

  1. … 2 more files in changeset.
LDT-1903 Air Trap adding duty cycle out of range alarm