Update alarms and Input/Output sections.

Update alarms and Input/Output sections.

Also, add option individual parameter override for substitutions pump rate as well.

Also, add option individual parameter override for substitutions pump rate as well.

Space after parenthesis. Please add the same in else portion as well (when treatment is in progress, update the rate).

Space after parenthesis. Please add the same in else portion as well (when treatment is in progress, update the rate).

fixed.

fixed.

fixed

fixed

fixed

fixed

fixed

fixed

added

added

This needs TD source update and also dialin update, keep informed system team about this change as this would impact DD demo running.

This needs TD source update and also dialin update, keep informed system team about this change as this would impact DD demo running.

Please add HDF Pump speed register and add a monitor in the substitution pump controller to compare against the target speed and if the difference is out of range, trigger an alarm.

Please add HDF Pump speed register and add a monitor in the substitution pump controller to compare against the target speed and if the difference is out of range, trigger an alarm.

Substitution pump

Substitution pump

Missing comment.

Missing comment.

update comment. next one as well.

update comment. next one as well.

blank line after the banner and maintain alphabetical order of include files.

blank line after the banner and maintain alphabetical order of include files.

LEAHI-DD-FIRMWARE-LDT-4001_HDF - FW Implementation - 2/4:
LEAHI-DD-FIRMWARE-LDT-4001_HDF - FW Implementation - 2/4:
LEAHI-DD-FIRMWARE-LDT-3975_Water Treatment and Degassing - FW Iteration - 4/4:
LEAHI-DD-FIRMWARE-LDT-3975_Water Treatment and Degassing - FW Iteration - 4/4:
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.
On DD f/w side, this should just be a simple pump controller. States should be Off and On (control to commanded Qsub). There is no flow sensor for feedback control, so assuming this is open loop si...

On DD f/w side, this should just be a simple pump controller. States should be Off and On (control to commanded Qsub). There is no flow sensor for feedback control, so assuming this is open loop similar to UF pump.
GenDialysate mode will have an HDF sub-state that signals this controller from Off to On and would also call separate controller for heater and separate (Switches) monitor for door sensors.

With all of that in mind, I think this unit should have a simpler name like SubPump.

I think there should be some state monitoring for potential alarm(s) in here too. Similar to BP door on TD f/w side where the current mode/state can signal this unit that it expects specific switch...

I think there should be some state monitoring for potential alarm(s) in here too. Similar to BP door on TD f/w side where the current mode/state can signal this unit that it expects specific switch state(s) (e.g. while in HDF state, we would expect the cap to be open and stowed and we would want to alarm if that was not the case).

Remove extra blank line.

Remove extra blank line.

LDT-3272 addressed the code review comments

  1. … 2 more files in changeset.
Where is call to HDF controller?

Where is call to HDF controller?

Monitors should come before mode mgmt above.

Monitors should come before mode mgmt above.

What about HemoDiaFiltration.h?

What about HemoDiaFiltration.h?

What about initHDF?

What about initHDF?

Where is HemoDiaFiltration.h?

Where is HemoDiaFiltration.h?

Why does this file name have pound signs pre/post?

Why does this file name have pound signs pre/post?