Not needed? Add another BOOL for disinfection option received.

Not needed? Add another BOOL for disinfection option received.

Initialize this.

Initialize this.

Remove TODO

Remove TODO

Remove disable of ejector motor.

Remove disable of ejector motor.

Remove RotaryValve.h. It's obsolete and we should be using higher level controller Valves.h anyway.

Remove RotaryValve.h. It's obsolete and we should be using higher level controller Valves.h anyway.

Just move msg handlers from Messaging to here so you don't need a signal.

Just move msg handlers from Messaging to here so you don't need a signal.

Move to .c file to make private.

Move to .c file to make private.

Not needed.

Not needed.

Do we need summary review msgs? 0xA9 and 0xAA cover request for summary data.

Do we need summary review msgs? 0xA9 and 0xAA cover request for summary data.

Missing heat citric

Missing heat citric

Put back to reserved

Put back to reserved

LEAHI-TD-FIRMWARE-LDT-4050_Post Treatment - FW Implementation - 1/3:
LEAHI-TD-FIRMWARE-LDT-4050_Post Treatment - FW Implementation - 1/3:
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 14 more files in changeset.
We can rename priorState to previousInstallState. Also, it should be declared as a static variable to be able to retain previousState value when the control goes out of the function.

We can rename priorState to previousInstallState.
Also, it should be declared as a static variable to be able to retain previousState value when the control goes out of the function.

There is still merge conflict? Please resolve merge conflict and build the code in your local before pushing.

There is still merge conflict?
Please resolve merge conflict and build the code in your local before pushing.

This merge conflict is still unresolved.

This merge conflict is still unresolved.

leak in the comment section.

leak in the comment section.

Where is alarm 191..203?

Where is alarm 191..203?

please add bpLeftHomeTimerCounter to inputs and outputs.

please add bpLeftHomeTimerCounter to inputs and outputs.

please add bpLeftHomeTimerCounter to inputs and outputs.

please add bpLeftHomeTimerCounter to inputs and outputs.

We only add File level or global variables in Inputs and Outputs. Please remove functions from the input section of this and other functions.

We only add File level or global variables in Inputs and Outputs.
Please remove functions from the input section of this and other functions.

Treatment mode will also need one of these (later). Should we add PreTx to the function name to differentiate?

Treatment mode will also need one of these (later). Should we add PreTx to the function name to differentiate?

Add a detail for Message Sent for the broadcast data message.

Add a detail for Message Sent for the broadcast data message.

create local Bool

create local Bool

Add a detail for Message Sent for the broadcast data message.

Add a detail for Message Sent for the broadcast data message.

There should be a \b before the type of details update it for all

There should be a \b before the type of details
update it for all

please add a comment.

please add a comment.