HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi
DENBUG-328 changed the get raw air trap function to get air trap function for alarm 215. Added air trap state change events

HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
RESOLVED IN CODEWALKTHROUGH

RESOLVED IN CODEWALKTHROUGH

DENBUG-328 addressed the code review comments

Done

Done

Please add the comments mentioning the reason/purpose for the checking fluid level timeout alarm?

Please add the comments mentioning the reason/purpose for the checking fluid level timeout alarm?

DENBUG-328 updated the air trap upper level fluid detection code

DENBUG-328 added persistent alarm for the air trap upper level fluid detection

HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 116 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

Delete the #define if not using it anymore.

Delete the #define if not using it anymore.

DENBUG-285 fixed the VBT in air trap

DENBUG-285 removed the VBV code in air trap

DENBUG-285 added code to set the VBV to close in air trap open state

DENBUG-285 delayed closing VBT 50ms

DENBUG-285 added code to close VBV when filling the air trap