LEAHI-DD-FIRMWARE-LDT-3980

control interval is different. Control interval must be separated else we don't get finer control on bicarb conductivity. also the dead band is different. need to figure out whether we can combine ...

control interval is different. Control interval must be separated else we don't get finer control on bicarb conductivity. also the dead band is different. need to figure out whether we can combine or not

This code looks very similar to bicarb code above. Do we really need separate handling?

This code looks very similar to bicarb code above. Do we really need separate handling?

Remove extra blank line.

Remove extra blank line.

Add unit of measure to definition name (e.g. _PSI). Should be done for all definitions w/ a unit of measure to prevent unit errors.

Add unit of measure to definition name (e.g. _PSI). Should be done for all definitions w/ a unit of measure to prevent unit errors.

LEAHI-DD-FIRMWARE-LDT-3980_Dialysate Composition - FW Implementation (Dry Bicarbonate) - 1/1:
LEAHI-DD-FIRMWARE-LDT-3980_Dialysate Composition - FW Implementation (Dry Bicarbonate) - 1/1: