LEAHI-TD-FIRMWARE-LDT-2240

I would keep blank line between library includes and regular includes.

I would keep blank line between library includes and regular includes.

We can't have 2 alarms with ID 196. The enum only has the empty saline bag alarm. Looks like alarm 197 is duplicate of this 196 alarm you've added. I think you should just remove this new alarm 196...

We can't have 2 alarms with ID 196. The enum only has the empty saline bag alarm. Looks like alarm 197 is duplicate of this 196 alarm you've added.
I think you should just remove this new alarm 196 that you've added here.

Fix merge conflict stuff here.

Fix merge conflict stuff here.

Fix merge conflict stuff here.

Fix merge conflict stuff here.

Bubble status, Low and high level level sensors.

Bubble status, Low and high level level sensors.

Input : Blood pump door status?

Input : Blood pump door status?

To be restored?

To be restored?

Are these being initialized in the init function?

Are these being initialized in the init function?

Need to resolve the merge issues?

Need to resolve the merge issues?

It looks like you have this as a T/F response from UI. UI should not be determining whether the modality matches - that's our job. UI should be telling us which type of tubing set was installed (HD...

It looks like you have this as a T/F response from UI. UI should not be determining whether the modality matches - that's our job. UI should be telling us which type of tubing set was installed (HD or HDF type). Then when we get this response msg from UI, we will compare the type to the user selected modality (from TxParams) to see if it matches.

Add doxygen comments to the right of each structure field.

Add doxygen comments to the right of each structure field.

Why is this commented out? We will get stuck in this state if we don't move on to the pressure leak test states.

Why is this commented out? We will get stuck in this state if we don't move on to the pressure leak test states.

For now, we need to set authResponseReceived and authResponseValidTubingSet and authResponseModalityAccepted to TRUE until UI is properly responding to authentication request (TODO).

For now, we need to set authResponseReceived and authResponseValidTubingSet and authResponseModalityAccepted to TRUE until UI is properly responding to authentication request (TODO).

For now, we need to set barcodeScanResponseReceived to TRUE here until barcode scan command is implemented.

For now, we need to set barcodeScanResponseReceived to TRUE here until barcode scan command is implemented.

Add blank line before return statement.

Add blank line before return statement.

Keep this blank line above group header.

Keep this blank line above group header.

Something is wrong here. We already had reg 33 at line 192 above. Reconcile with HDD.

Something is wrong here. We already had reg 33 at line 192 above. Reconcile with HDD.

These alarms are in the wrong place. Their enums are 187..199. So they should be at the very end of this table, not here near the beginning.

These alarms are in the wrong place. Their enums are 187..199. So they should be at the very end of this table, not here near the beginning.

Remove this blank line.

Remove this blank line.

I missed updating this earlier, I’ll push the updated code.

I missed updating this earlier, I’ll push the updated code.

Can't have 3 values. Maximum 2 values allowed.

Can't have 3 values. Maximum 2 values allowed.

Removedd

Removedd

added

added

Updated

Updated

Remove start state.

Remove start state.

Updated

Updated

updated

updated

Yes, removed

Yes, removed

Removed

Removed

yes, removed

yes, removed