components

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-13743: IMP: Mostly worked on Alarm Dialog/Bar

- Changed the Alarm Dialog (Title decreased by 4, Description increased by 4) / Bar (increased by 6) font size,

- Adjusted the AlarmDialog Timer position and layout.

- Removed the mouseArea from the WaitDone component.

- Added titel/description fontPixedSize for the NotificationDialog and NotificationBar

- Clear the reasontext on home screen when is not visible anymore.

- Clear the reasontext on SettingsDG when is not visible anymore.

- Used NotificationBarSmall instead of NotificationBar in:

- ModalDialog

- MainHome

- TreatmentAdjustmentBase

- TreatmentFlowBase

  1. … 12 more files in changeset.
DEN-13645 IMP: DEN-12958: UI shall generate the POST fail alarm if the SD-Card not present and stop logging

__________

- Log Export

- Added the Alarm generation on the GuiController if SD-Card is not ready, or is/becomes read-only.

- AlarmGenerator

- Added the SD-Card failure alarm - ALARM_ID_UI_POST_FAILURE_SDCARD

- Added alarm clear - ALARM_ID_NO_ALARM

- Added a flag to stop logging if the SD-Card is not ready or is readonly.

- Removed pause on log clean up.

- Improved the DeviceController SD-Card, USB status detection.

- Fixed and improved the MVC connection betweein device/Application/Gui controllers.

- A CLI switch ( -S ) added to disable the feature that is disabling the log if the SD-Card doesn't exist. So if the -S switch is used while calling the denali application, the logging will NOT be disabled, if SD-Card fails.

- In the ExportButton and USBButton(eject) removed the enable logic and has to be done outside of the component. It has been done for the Settings/export and Post-Tx/Review.

- Added title and file count for each SD-Card and USB-Drive panels.

__________

- Run.sh

- Added a check for the CloudSync and if the folder doesn't exist it won't do the rest

- Kill the denali application, in case it is still running.

- Added SD-Card type as part of its detection and if it is not ext4 format it fails the test.

- Optimized the mount point detection.

__________

- NoCANBus

- added the NoCANBus to the fault and init modes to not to navigate to those if NoCANBus is true. Otherwise it conflict with navigate to the home.

- Added a log in case the NoCANBus is enabled.

__________

- Settings

- Converted the '\n' in the settings values to actual newline character.

__________

- Thread:

- Moved the QuiController in its own thread, _Gui_Thread.

__________

- Alarms

- Added an emit to update the Alarm Dialog with the latest title/message, received from Alarms.conf In case there was an alarm before even Alarm.conf is read.

  1. … 26 more files in changeset.
DEN-13645: IMP: General Implementation

DEN-12836: Treatment Report/log export modifications

- Finalizing the log export.

  1. … 3 more files in changeset.
DEN-13645: IMP: General Implementation

DEN-12836: Treatment Report/log export modifications

- Update the eject and export button status for the export service and treatment

  1. … 2 more files in changeset.
DEN-13489 IMP: General Implementation

DEN-13035: Update alarm body messages

- Removed the exclamation mark icon form the Alarm Dialog.

- Removed the exclamation mark icon form the Alarm Bar.

- Made the whole AlarmDialog title bar doubleClick and diagnostics opener if the alarm is not minimizable.

- Moved the Alarm ID from title to the bottom left of the dialog with format: "ID:<id>".

- Clipped the AlarmDialon body message area, crop the text if longer than dialog area.

- Changed the "settings" property name of the VSettings to "instructions".

- Changed the date/time formats keys to simplify the code and conf file.

- Implemented the Alarm Body messages, to load from Alams/Alarms.conf, instead of the FW.

- The FW texts/documents will still being used if the Alarms.conf doesn't have a definition.

- A instruction base capability has also been added to the MVC, but UI still not supporting it.

  1. … 18 more files in changeset.
DEN-13049: UI Treatment Parameters Entry: Some values for treatment parameter entry on treatment parameter entry screen in create treatment do not match SRS/PRS

- A bug found on multiple, ON,OFF toggle, slider sends minimum instead of '0' while ON.

- Slider improved.

- Added float rounding of the request message logs to 3.

  1. … 4 more files in changeset.
DEN-9129: Unable to proceed past "Create a Custom Treatment" when Arterial and Venous Pressure (High) Limits maximum values are initially selected

Also modified the SliderCreateTreatment.qml to show by the default the zero value not the OFF since it is only two Heparin Bolus/Dispensing which display OFF for zero value.

  1. … 2 more files in changeset.
DEN-13049: UI Treatment Parameters Entry: Some values for treatment parameter entry on treatment parameter entry screen in create treatment do not match SRS/PRS

  1. … 4 more files in changeset.
RESOLVED.

RESOLVED.

Sure Sean, if that is the decision, will do it soon and create a case for that. I also believe some other CL Switches need to be removed, which were only implemented for our testing at the time (CA...

Sure Sean, if that is the decision, will do it soon and create a case for that.
I also believe some other CL Switches need to be removed, which were only implemented for our testing at the time (CANBus).

Issue DEN-13441 - MR: Make the Dialin log permanent with no switch [CL switches review/removal] has been successfully created

If this is to log Dialin messages, we should always do that (no switch). If a Dialin command is given/responded to, we definitely want to know about that - so want to see it in the log. Dialin comm...

If this is to log Dialin messages, we should always do that (no switch). If a Dialin command is given/responded to, we definitely want to know about that - so want to see it in the log. Dialin commands are always one-shot type messages, so will not impact log size.

RESOLVED.

RESOLVED.

RESOLVED.

RESOLVED.

RESOLVED

RESOLVED

This code is going to be used later in the Log Export for slow transfer & confirmation for the USB port slow transfer issue. I kept it in the application repo to keep it safe. Commented out to not ...

This code is going to be used later in the Log Export for slow transfer & confirmation for the USB port slow transfer issue.
I kept it in the application repo to keep it safe.
Commented out to not affect the coverage.

http://dvm-linux02:8060/cru/#UI-DEN-12716-2CFR-30978 in PostTreatmentReview.qml@40

http://dvm-linux02:8060/cru/#UI-DEN-12716-2CFR-30978
in PostTreatmentReview.qml@40

Where would this value change?

Where would this value change?

At the moment I added that switch I assumed it should only be used for V&V and SysV testing.

At the moment I added that switch I assumed it should only be used for V&V and SysV testing.

This is the requirement. PRS 417: The HD Device manufacturing default timezone shall be (UTC-08:00) Pacific Time.

This is the requirement.
PRS 417: The HD Device manufacturing default timezone shall be (UTC-08:00) Pacific Time.

Why is main.cpp all commented out?

Why is main.cpp all commented out?

All systems will be setup for Pacific time zone?

All systems will be setup for Pacific time zone?

Why must this be removed later?

Why must this be removed later?

UI-DEN-12716_UI BN S70
UI-DEN-12716_UI BN S70
DEN-13127: LTS: Development Support

- [Export Button Enabling in VM #61|https://github.com/Diality/testsuites/issues/61]

  1. … 8 more files in changeset.
DEN-13134: IMP: General Implementation for any bugs and any other implementations

DEN-12836: Treatment Report/log export modifications

- Improved the SDItem and renamed to SDCProgressItem.

- Added another separate progress circle component for the USB Driver [USBProgressItem]

- Adding Space check for USB as well to get the USBProgressCircle update.

- Optimized ExportButton and added two size for Settings screen(normal) and Treatment Review screen(small)

- Improved USB button comonent

- Added a global byte size converter in Variable.qml

- Improved MainMenu component

- Added a notification for the TreatmentFlowBase.qml main component.

- Used the notification of the TreatmentFlowBase.qml for the Treatment Review Screen Log export to display feedback to the user.

- Added an individual screen for the log export.

- Removed the Export Log and USB eject from the main settings option screen.

- Added an option for the Export log on the settings home screen which is currently enabled for non-service users too.

- Fixed the DGVersionResponse, log message.

- Put a comment in the FileHandler.cpp about the Qt API QStorageInfo::mountedVolume() which seems so slow on detecting mounted volumes when a USB derive is just plugged in (It could be system issue, since it has also been observed from shell console.)

- Added error handling to the Treatment Log/Report export

DEN-12726: Denali application crashing

- Modified the Logger Maxusage limit to keep 50% of the data/event log and 15% of the service log and keep 35% free, after the log clean up triggered by 15% free storage detected.

DEN-13051: Encounter alarm 6 and SD card detection problem after updating UI version

DEN-13156: Settings screen menu should disappear on service mode

- Improved the MainStack.qml navigation rules.

- Improved the Patient ID keyboard handling, to hide the keyboard, when not on the screen.

- Improved the SettingsStack.qml navigation rules.

DEN-13128: KBM: Cloud Sync Integration

- Saving the CS log in UTC timezone.

DEN-13127: LTS: Development Support

- Fixed the issue found on the End-Treatment screen Confirm button which didn't respect the last instruction page enable only rule.

  1. … 40 more files in changeset.
DEN-12558: HD does not transition to service mode

- implemented the service mode confirmation

- decreased the dialog animation duration

- renamed the service dates MVC to differentiate with service mode.

  1. … 32 more files in changeset.
DEN-12541: Get the SD-Card size from the system

  1. … 4 more files in changeset.