DEN-14955 : Quick Start - Main treatment Screens - Update screens to appropriately use or not use rounded sliders

  1. … 18 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 11 more files in changeset.
Merge branch 'DEN-14909-UI-VD-S92-QS-Sliders' into staging

  1. … 6 more files in changeset.
DEN-14912 Quick Start : Sliders - Remove Component and Loaders from TickMarks and changed to Rectangle - Made additional fixes based on CR comments

  1. … 5 more files in changeset.
DEN-14912 Quick Start: Arrows - Additional changes based on Behrouz's feedback

  1. … 4 more files in changeset.
DEN-14912 Quick Start - Arrows - Modify code based on feedback from Behrouz -- take out the CircleButton and make use of TouchRect -- Add left and right arrow images

  1. … 6 more files in changeset.
DEN-14912: Quick Start - Arrows - Add implementation for value increment/decrement buttons

  1. … 6 more files in changeset.
DEN-14912: Quick Start Sliders -Improvement to the rangeSlider's look and feel for quick start

  1. … 4 more files in changeset.
DEN-14912: Quick Start - Sliders - Improvements to the sliders look and feel for the quick start

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 15 more files in changeset.
DEN-13672 UI : Saline Bolus Section on Main Treatment Screen Does Not Match SRSUI/PRS

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 608 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 34 more files in changeset.
DEN-13962: UI: Update pressure limits to match treatment params and UI screen

- removed the QML duplicate definitions of arterial/venous

- added monitoring values and used them in monitoring components of arterial/venous.

  1. … 17 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 28 more files in changeset.
Merge branch 'DEN-13965-UI-BN-S80' into staging

  1. … 12 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 95 more files in changeset.
Merge branch 'DEN-13198-UI_BN_75' into staging

  1. … 45 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 100 more files in changeset.
Merge branch 'DEN-12640-UI-BN-S79' into develop

DEN-12640: UI: Pressure readings on blood prime / rinseback

- The In-Line Blood Pressure monitor (ILBPM) has been added to the Rinseback and blood prime screen.

- It has been requested to place the ILBPM, at same place on all three screens (Main Treatment is the third one), but on the rinseback screen we don't have space for that. So it has been moved to top/left (it was buttom/left) and swapped with the FLOWS section.

- It needs to be negotiated with the team to make sure everyone agrees.

  1. … 4 more files in changeset.
DEN-13049: UI Treatment Parameters Entry: Some values for treatment parameter entry on treatment parameter entry screen in create treatment do not match SRS/PRS

  1. … 5 more files in changeset.
DEN-13134: IMP: General Implementation for any bugs and any other implementations

DEN-12836: Treatment Report/log export modifications

- Improved the SDItem and renamed to SDCProgressItem.

- Added another separate progress circle component for the USB Driver [USBProgressItem]

- Adding Space check for USB as well to get the USBProgressCircle update.

- Optimized ExportButton and added two size for Settings screen(normal) and Treatment Review screen(small)

- Improved USB button comonent

- Added a global byte size converter in Variable.qml

- Improved MainMenu component

- Added a notification for the TreatmentFlowBase.qml main component.

- Used the notification of the TreatmentFlowBase.qml for the Treatment Review Screen Log export to display feedback to the user.

- Added an individual screen for the log export.

- Removed the Export Log and USB eject from the main settings option screen.

- Added an option for the Export log on the settings home screen which is currently enabled for non-service users too.

- Fixed the DGVersionResponse, log message.

- Put a comment in the FileHandler.cpp about the Qt API QStorageInfo::mountedVolume() which seems so slow on detecting mounted volumes when a USB derive is just plugged in (It could be system issue, since it has also been observed from shell console.)

- Added error handling to the Treatment Log/Report export

DEN-12726: Denali application crashing

- Modified the Logger Maxusage limit to keep 50% of the data/event log and 15% of the service log and keep 35% free, after the log clean up triggered by 15% free storage detected.

DEN-13051: Encounter alarm 6 and SD card detection problem after updating UI version

DEN-13156: Settings screen menu should disappear on service mode

- Improved the MainStack.qml navigation rules.

- Improved the Patient ID keyboard handling, to hide the keyboard, when not on the screen.

- Improved the SettingsStack.qml navigation rules.

DEN-13128: KBM: Cloud Sync Integration

- Saving the CS log in UTC timezone.

DEN-13127: LTS: Development Support

- Fixed the issue found on the End-Treatment screen Confirm button which didn't respect the last instruction page enable only rule.

  1. … 44 more files in changeset.
wrapping up multiple sprint codes

- LOG: Added a CLI option to disable the Check-In logs by adding -k, --disable-check-in-log at the CLI. This was interrupting the log tests. The build macro has been removed.

- POST: During the coverage find out that the getter functions have not even been called in the ApplicationPost.

- CloudSync: Added the Treatment Code(TxCode), received from the cloud after the Treatment Report stored by Cloud System. It is a message communication between UI<=(2008)=CS.

- CloudSync: Two error added for the 2008: TxCode

- CloudSync: Due to some issues on Cloud System regarding the credential expiration, It has been asked to add a temporary workaround to send the credentials (1004) before message 1006:Device State, 1007:Treatment Report.

- DEN-12727: Treatment Parameter Screen: Added more space between the parameters components specially the Temperature slider and the option on top.

- DEN-12829: Enabled the back button on the Start Treatment Screen

- DEN-12761: Enabled an empty Patient ID.

It has not yet been implemented on the CloudSync App, so if sent empty CloudSystem will later reject to store the TxReport.

When later implemented an unassigned random Patient ID will be created for that TxReport to be later defined and assigned.

- DEN-12837, DEN-12840: The Disinfection Scheduling Entry and the Confirm button have been removed from the "Dialysate Generator Settings" screen.

- DEN-12837, DEN-12840: The "Dialysate Generator Settings" changed to "DG Cleaning" for the menu option and the screen title.

- Settings: The Clear Alarm Condition button has been removed from the settings screen for the release.

- Settings: The options which has no implementation for the phase 1 have been removed from the Settings screen Service mode options. [Language, Calibration, SW Update, Factory Reset]

- DEN-12841: Reworded the password entries in Service screen and for WiFi Join Screen.

- TxCode, TxReport: removed the commented out Treatment Report Parameters from the code.

- LOG: added one character space int he DEBUG log for the Not Enough Data in Types.

- Settings: The Service Dates default values changed from 0 to "" since they are strings.

- Settings: Added some clarification for the Home screen greetings

  1. … 32 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 77 more files in changeset.
DEN-12106: Heparin Delivery Active Text Does Not Appear

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 584 more files in changeset.
DEN-11814: Heparin Minimum Dispensing Parameters Should Be 0.2

- updated the min range regarding the recent PRS-347, PRS-348

- Fixed the minimum precision of the heparin.

  1. … 4 more files in changeset.
DEN-8145: Issues with UI Create Treatment screen

- Restricted the Patient ID entry to be max 20 characters and has to have at least 1 alphanumeric character without space, regarding the conversation with KeborMed.

  1. … 2 more files in changeset.
DEN-9142: Meetings & Support - General Support

- Updating the blood/dialysate flow ranges/resolution.

- Updated the Serial messages length from 10 to 15 bytes.

- Having some work done testing the plotting component.

  1. … 7 more files in changeset.