PreTreatmentCreateStack.qml

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
CS-13: Patient ID set as fix if not entered by user

- Enabled the CS timers to look up for the pending txr.

- Fix a typo which caused Settings screens notification cover by Minimized Alarm.

- Added a default Patient ID "NoPatientIDEntered" to be stored as, in the Txr, in case skipped.

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 612 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 5 more files in changeset.
DEN-16637, DEN-16640, DEN-16641 : P1_Staging Dry-Run found issues.

DEN-16637: SW: Top Alarm on Debug Screen Does Not Change When Alarm is Cleared

DEN-16640: SW: Blood Measurement Interval Set to OFF During Treatment When Set to 5 min in Pre-treatment

DEN-16641: UI: Heparin Syringe Installation Screen Missing

  1. … 4 more files in changeset.
DEN-16637, DEN-16640, DEN-16641 : P1_Staging Dry-Run found issues.

DEN-16637: SW: Top Alarm on Debug Screen Does Not Change When Alarm is Cleared

DEN-16640: SW: Blood Measurement Interval Set to OFF During Treatment When Set to 5 min in Pre-treatment

DEN-16641: UI: Heparin Syringe Installation Screen Missing

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 612 more files in changeset.
DENBUG-75: UI: Treatment log data via USB does not match with data displayed on the Treatment Review Screen

CS-5: @ UI: TxCode generation on device

~ Added QRCode for the TxCode

- Added a Third-Party Open-Source MIT Licence QRCode generator from the following link:

* Copyright (c) Project Nayuki. (MIT License)

* https://www.nayuki.io/page/qr-code-generator-library

- Created a QRCode component

- Added the QRCode generation functionality in the GuiView

- Added QRCode for both the Pre-Tx-Confirmation (UI Generated) and Post-Tx-Review (CS received)

- Moved the CS TxCode Receive wait "..." from C++ to QML to keep the TxCode empty while wait to avoid displaying the "..." as a QRCode.

- Animated the CS TxCode Receive wait "..."

~ Added development testability for the CloudSync communication

~ Fixed the date change buf file watch to wait for the CS to get the checkin and create the out buf file.

CS-3: CS: Heparin OFF o the HCP portal Treatment report

- Fixed some issues in the create treatment screen on heparin type

- Combined the separated values for log and GUI into one values for both.

- Added the functionality for the TreatmentConfirm (Pre-Tx), TreatmentReview (Ppost-Tx), TreatmentReport (log) for the heparin values.

DENBUG-75: UI: Treatment log data via USB does not match with data displayed on the Treatment Review Screen

DENBUG-75: UI: Treatment log data via USB does not match with data displayed on the Treatment Review Screen

- removed the qtcode component from the two screen of the treatment confirm, review for P1 release.

  1. … 19 more files in changeset.
DENBUG-207: UI: UI does not reset to default parameters upon new patient

DEN-12728: UI: UI does not clean up Treatment Parameter settings for different patient

- the reset is added to the treatment parameters

- the foundation of the pre-population has been added but still there is parts need to be added to store the sliders' On/Off switch to be restored which is needed for the slider to set properly.

  1. … 4 more files in changeset.
DEN-12728: UI: UI does not clean up Treatment Parameter settings for different patient

- the reset is added to the treatment parameters

- the foundation of the pre-population has been added but still there is parts need to be added to store the sliders' On/Off switch to be restored which is needed for the slider to set properly.

  1. … 4 more files in changeset.
CS-3: CS: Heparin OFF o the HCP portal Treatment report

- Fixed some issues in the create treatment screen on heparin type

- Combined the separated values for log and GUI into one values for both.

- Added the functionality for the TreatmentConfirm (Pre-Tx), TreatmentReview (Ppost-Tx), TreatmentReport (log) for the heparin values.

  1. … 7 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 20 more files in changeset.
DEN-15911: Post-treatment screen displays heparin type and heparin concentration even if heparin rate and dispensing were disabled in pre-treatment

- Separated the Log and the Gui values for now to address the cloud log later, when the cloud is ready to not to reject Logs to have NONE.

  1. … 8 more files in changeset.
DEN-12842: Added - UI: prepopulate option on the create treatment screen (Phase 1b)

Refactored the adjust parameters model to add a stored prepopulate model

Added resetting and setting selection to gridSelection.qml component

Add a set of properties to store pre-confirmed treatment parameters

Add a model for storing unconfirmed data

Add handling of resetting and pre-populating treatment parameters

  1. … 7 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
DEN-15493 : Usability: UI, FW: rinseback flow rate

- removed the rinseback range slider form the create treatment screen and passing the default value of the 200 set in the configurations for the message.

- min max has also changed to 100 - 300 in ui.config repository.

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 54 more files in changeset.
DEN-15359: Usability: Arterial and Venous Pressure Alarm Limits

- DEN-15387: UI: Updates for treatment parameter entry (A/V alarm window)

- DEN-15388: UI: Updates for in-treatment adjustment of alarm parameter window (pressure alarm screen)

- Monitoring components for the TxHome, Rinseback, BloodPrime

  1. … 35 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 23 more files in changeset.
DEN-13537 UI: Wi-Fi Network Settings not all preserved by Device after reboot

- Decided to disable the WiFi static IP settings fro phase 1.

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 9 more files in changeset.
DEN-14955 Tx Create Reject

- code enhancement

- option buttons rejection improvement

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 608 more files in changeset.
DEN-14391 , DEN-14393 : Qt 5.15.10 [QML run-time issues]

- Fixed all the QML warning and errors

- Enabled the Clang-Tidy, Clazy and cppcheck and fixed some.

- Removed unused images

- made the Denali POST logo transparent.

  1. … 56 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 95 more files in changeset.
Merge branch 'DEN-13198-UI_BN_75' into staging

  1. … 45 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 100 more files in changeset.
Fixed the issue on the PatientID confirm text

  1. … 1 more file in changeset.
DEN-13134: IMP: General Implementation for any bugs and any other implementations

DEN-12836: Treatment Report/log export modifications

- Improved the SDItem and renamed to SDCProgressItem.

- Added another separate progress circle component for the USB Driver [USBProgressItem]

- Adding Space check for USB as well to get the USBProgressCircle update.

- Optimized ExportButton and added two size for Settings screen(normal) and Treatment Review screen(small)

- Improved USB button comonent

- Added a global byte size converter in Variable.qml

- Improved MainMenu component

- Added a notification for the TreatmentFlowBase.qml main component.

- Used the notification of the TreatmentFlowBase.qml for the Treatment Review Screen Log export to display feedback to the user.

- Added an individual screen for the log export.

- Removed the Export Log and USB eject from the main settings option screen.

- Added an option for the Export log on the settings home screen which is currently enabled for non-service users too.

- Fixed the DGVersionResponse, log message.

- Put a comment in the FileHandler.cpp about the Qt API QStorageInfo::mountedVolume() which seems so slow on detecting mounted volumes when a USB derive is just plugged in (It could be system issue, since it has also been observed from shell console.)

- Added error handling to the Treatment Log/Report export

DEN-12726: Denali application crashing

- Modified the Logger Maxusage limit to keep 50% of the data/event log and 15% of the service log and keep 35% free, after the log clean up triggered by 15% free storage detected.

DEN-13051: Encounter alarm 6 and SD card detection problem after updating UI version

DEN-13156: Settings screen menu should disappear on service mode

- Improved the MainStack.qml navigation rules.

- Improved the Patient ID keyboard handling, to hide the keyboard, when not on the screen.

- Improved the SettingsStack.qml navigation rules.

DEN-13128: KBM: Cloud Sync Integration

- Saving the CS log in UTC timezone.

DEN-13127: LTS: Development Support

- Fixed the issue found on the End-Treatment screen Confirm button which didn't respect the last instruction page enable only rule.

  1. … 44 more files in changeset.
wrapping up multiple sprint codes

- LOG: Added a CLI option to disable the Check-In logs by adding -k, --disable-check-in-log at the CLI. This was interrupting the log tests. The build macro has been removed.

- POST: During the coverage find out that the getter functions have not even been called in the ApplicationPost.

- CloudSync: Added the Treatment Code(TxCode), received from the cloud after the Treatment Report stored by Cloud System. It is a message communication between UI<=(2008)=CS.

- CloudSync: Two error added for the 2008: TxCode

- CloudSync: Due to some issues on Cloud System regarding the credential expiration, It has been asked to add a temporary workaround to send the credentials (1004) before message 1006:Device State, 1007:Treatment Report.

- DEN-12727: Treatment Parameter Screen: Added more space between the parameters components specially the Temperature slider and the option on top.

- DEN-12829: Enabled the back button on the Start Treatment Screen

- DEN-12761: Enabled an empty Patient ID.

It has not yet been implemented on the CloudSync App, so if sent empty CloudSystem will later reject to store the TxReport.

When later implemented an unassigned random Patient ID will be created for that TxReport to be later defined and assigned.

- DEN-12837, DEN-12840: The Disinfection Scheduling Entry and the Confirm button have been removed from the "Dialysate Generator Settings" screen.

- DEN-12837, DEN-12840: The "Dialysate Generator Settings" changed to "DG Cleaning" for the menu option and the screen title.

- Settings: The Clear Alarm Condition button has been removed from the settings screen for the release.

- Settings: The options which has no implementation for the phase 1 have been removed from the Settings screen Service mode options. [Language, Calibration, SW Update, Factory Reset]

- DEN-12841: Reworded the password entries in Service screen and for WiFi Join Screen.

- TxCode, TxReport: removed the commented out Treatment Report Parameters from the code.

- LOG: added one character space int he DEBUG log for the Not Enough Data in Types.

- Settings: The Service Dates default values changed from 0 to "" since they are strings.

- Settings: Added some clarification for the Home screen greetings

  1. … 32 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 584 more files in changeset.