Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 9 more files in changeset.
CS-13: Patient ID set as fix if not entered by user

- Enabled the CS timers to look up for the pending txr.

- Fix a typo which caused Settings screens notification cover by Minimized Alarm.

- Added a default Patient ID "NoPatientIDEntered" to be stored as, in the Txr, in case skipped.

  1. … 5 more files in changeset.
CS-13: Patient ID set as fix if not entered by user

- Enabled the CS timers to look up for the pending txr.

- Fix a typo which caused Settings screens notification cover by Minimized Alarm.

- Added a default Patient ID "NoPatientIDEntered" to be stored as, in the Txr, in case skipped.

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 612 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 5 more files in changeset.
DEN-16637, DEN-16640, DEN-16641 : P1_Staging Dry-Run found issues.

DEN-16637: SW: Top Alarm on Debug Screen Does Not Change When Alarm is Cleared

DEN-16640: SW: Blood Measurement Interval Set to OFF During Treatment When Set to 5 min in Pre-treatment

DEN-16641: UI: Heparin Syringe Installation Screen Missing

  1. … 4 more files in changeset.
DEN-16637, DEN-16640, DEN-16641 : P1_Staging Dry-Run found issues.

DEN-16637: SW: Top Alarm on Debug Screen Does Not Change When Alarm is Cleared

DEN-16640: SW: Blood Measurement Interval Set to OFF During Treatment When Set to 5 min in Pre-treatment

DEN-16641: UI: Heparin Syringe Installation Screen Missing

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 612 more files in changeset.
DENBUG-75: UI: Treatment log data via USB does not match with data displayed on the Treatment Review Screen

CS-5: @ UI: TxCode generation on device

~ Added QRCode for the TxCode

- Added a Third-Party Open-Source MIT Licence QRCode generator from the following link:

* Copyright (c) Project Nayuki. (MIT License)

* https://www.nayuki.io/page/qr-code-generator-library

- Created a QRCode component

- Added the QRCode generation functionality in the GuiView

- Added QRCode for both the Pre-Tx-Confirmation (UI Generated) and Post-Tx-Review (CS received)

- Moved the CS TxCode Receive wait "..." from C++ to QML to keep the TxCode empty while wait to avoid displaying the "..." as a QRCode.

- Animated the CS TxCode Receive wait "..."

~ Added development testability for the CloudSync communication

~ Fixed the date change buf file watch to wait for the CS to get the checkin and create the out buf file.

CS-3: CS: Heparin OFF o the HCP portal Treatment report

- Fixed some issues in the create treatment screen on heparin type

- Combined the separated values for log and GUI into one values for both.

- Added the functionality for the TreatmentConfirm (Pre-Tx), TreatmentReview (Ppost-Tx), TreatmentReport (log) for the heparin values.

DENBUG-75: UI: Treatment log data via USB does not match with data displayed on the Treatment Review Screen

DENBUG-75: UI: Treatment log data via USB does not match with data displayed on the Treatment Review Screen

- removed the qtcode component from the two screen of the treatment confirm, review for P1 release.

  1. … 19 more files in changeset.
DENBUG-207: UI: UI does not reset to default parameters upon new patient

DEN-12728: UI: UI does not clean up Treatment Parameter settings for different patient

- the reset is added to the treatment parameters

- the foundation of the pre-population has been added but still there is parts need to be added to store the sliders' On/Off switch to be restored which is needed for the slider to set properly.

  1. … 4 more files in changeset.
DEN-12728: UI: UI does not clean up Treatment Parameter settings for different patient

- the reset is added to the treatment parameters

- the foundation of the pre-population has been added but still there is parts need to be added to store the sliders' On/Off switch to be restored which is needed for the slider to set properly.

  1. … 4 more files in changeset.
CS-3: CS: Heparin OFF o the HCP portal Treatment report

- Fixed some issues in the create treatment screen on heparin type

- Combined the separated values for log and GUI into one values for both.

- Added the functionality for the TreatmentConfirm (Pre-Tx), TreatmentReview (Ppost-Tx), TreatmentReport (log) for the heparin values.

  1. … 7 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 20 more files in changeset.
DEN-15911: Post-treatment screen displays heparin type and heparin concentration even if heparin rate and dispensing were disabled in pre-treatment

- Separated the Log and the Gui values for now to address the cloud log later, when the cloud is ready to not to reject Logs to have NONE.

  1. … 8 more files in changeset.
DEN-12842: Added - UI: prepopulate option on the create treatment screen (Phase 1b)

Refactored the adjust parameters model to add a stored prepopulate model

Added resetting and setting selection to gridSelection.qml component

Add a set of properties to store pre-confirmed treatment parameters

Add a model for storing unconfirmed data

Add handling of resetting and pre-populating treatment parameters

  1. … 7 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
DEN-15493 : Usability: UI, FW: rinseback flow rate

- removed the rinseback range slider form the create treatment screen and passing the default value of the 200 set in the configurations for the message.

- min max has also changed to 100 - 300 in ui.config repository.

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 54 more files in changeset.
DEN-15359: Usability: Arterial and Venous Pressure Alarm Limits

- DEN-15387: UI: Updates for treatment parameter entry (A/V alarm window)

- DEN-15388: UI: Updates for in-treatment adjustment of alarm parameter window (pressure alarm screen)

- Monitoring components for the TxHome, Rinseback, BloodPrime

  1. … 35 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 23 more files in changeset.
DEN-13537 UI: Wi-Fi Network Settings not all preserved by Device after reboot

- Decided to disable the WiFi static IP settings fro phase 1.

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 9 more files in changeset.
DEN-14955 Tx Create Reject

- code enhancement

- option buttons rejection improvement

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 608 more files in changeset.
DEN-14391 , DEN-14393 : Qt 5.15.10 [QML run-time issues]

- Fixed all the QML warning and errors

- Enabled the Clang-Tidy, Clazy and cppcheck and fixed some.

- Removed unused images

- made the Denali POST logo transparent.

  1. … 56 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 95 more files in changeset.
Merge branch 'DEN-13198-UI_BN_75' into staging

  1. … 45 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 100 more files in changeset.
Fixed the issue on the PatientID confirm text

  1. … 1 more file in changeset.
DEN-13134: IMP: General Implementation for any bugs and any other implementations

DEN-12836: Treatment Report/log export modifications

- Improved the SDItem and renamed to SDCProgressItem.

- Added another separate progress circle component for the USB Driver [USBProgressItem]

- Adding Space check for USB as well to get the USBProgressCircle update.

- Optimized ExportButton and added two size for Settings screen(normal) and Treatment Review screen(small)

- Improved USB button comonent

- Added a global byte size converter in Variable.qml

- Improved MainMenu component

- Added a notification for the TreatmentFlowBase.qml main component.

- Used the notification of the TreatmentFlowBase.qml for the Treatment Review Screen Log export to display feedback to the user.

- Added an individual screen for the log export.

- Removed the Export Log and USB eject from the main settings option screen.

- Added an option for the Export log on the settings home screen which is currently enabled for non-service users too.

- Fixed the DGVersionResponse, log message.

- Put a comment in the FileHandler.cpp about the Qt API QStorageInfo::mountedVolume() which seems so slow on detecting mounted volumes when a USB derive is just plugged in (It could be system issue, since it has also been observed from shell console.)

- Added error handling to the Treatment Log/Report export

DEN-12726: Denali application crashing

- Modified the Logger Maxusage limit to keep 50% of the data/event log and 15% of the service log and keep 35% free, after the log clean up triggered by 15% free storage detected.

DEN-13051: Encounter alarm 6 and SD card detection problem after updating UI version

DEN-13156: Settings screen menu should disappear on service mode

- Improved the MainStack.qml navigation rules.

- Improved the Patient ID keyboard handling, to hide the keyboard, when not on the screen.

- Improved the SettingsStack.qml navigation rules.

DEN-13128: KBM: Cloud Sync Integration

- Saving the CS log in UTC timezone.

DEN-13127: LTS: Development Support

- Fixed the issue found on the End-Treatment screen Confirm button which didn't respect the last instruction page enable only rule.