DEN-13049: UI Treatment Parameters Entry: Some values for treatment parameter entry on treatment parameter entry screen in create treatment do not match SRS/PRS

- A bug found on multiple, ON,OFF toggle, slider sends minimum instead of '0' while ON.

- Slider improved.

- Added float rounding of the request message logs to 3.

    • -3
    • +4
    ./components/SliderCreateTreatment.qml
  1. … 3 more files in changeset.
DEN-13049 - UI Treatment Parameters Entry: Some values for treatment parameter entry on treatment parameter entry screen in create treatment do not match SRS/PRS

- There was a bug in the fix which was related to the validation and caused to fail the treatment parameters validation.

- also removed the disable if the SD is not detected correctly. (QStorageInfo is not sending DeviceReady correctly)

- enabled the float rounding to 3 in the log

- Enabled the Create Treatment button rejection notification, since it seems it is not being handled by alarms as presumed.

- Fixed the BloodPressure interval value to display OFF when set to zero.

  1. … 2 more files in changeset.
DEN-9129: Unable to proceed past "Create a Custom Treatment" when Arterial and Venous Pressure (High) Limits maximum values are initially selected

Also modified the SliderCreateTreatment.qml to show by the default the zero value not the OFF since it is only two Heparin Bolus/Dispensing which display OFF for zero value.

    • -8
    • +4
    ./components/SliderCreateTreatment.qml
  1. … 1 more file in changeset.
DEN-13049: UI Treatment Parameters Entry: Some values for treatment parameter entry on treatment parameter entry screen in create treatment do not match SRS/PRS

    • -0
    • +1
    ./components/SliderCreateTreatment.qml
  1. … 2 more files in changeset.
DEN-13199: INT: DVT Support

- During DVT support found out the BCuff vitals debug item on the top of the screen is not working, and fixed ti by adding a real time version of the same actual vital properties which gets updated every time the vitals received.

  1. … 5 more files in changeset.
RESOLVED.

RESOLVED.

Sure Sean, if that is the decision, will do it soon and create a case for that. I also believe some other CL Switches need to be removed, which were only implemented for our testing at the time (CA...

Sure Sean, if that is the decision, will do it soon and create a case for that.
I also believe some other CL Switches need to be removed, which were only implemented for our testing at the time (CANBus).

Issue DEN-13441 - MR: Make the Dialin log permanent with no switch [CL switches review/removal] has been successfully created

If this is to log Dialin messages, we should always do that (no switch). If a Dialin command is given/responded to, we definitely want to know about that - so want to see it in the log. Dialin comm...

If this is to log Dialin messages, we should always do that (no switch). If a Dialin command is given/responded to, we definitely want to know about that - so want to see it in the log. Dialin commands are always one-shot type messages, so will not impact log size.

RESOLVED.

RESOLVED.

RESOLVED.

RESOLVED.

RESOLVED

RESOLVED

This code is going to be used later in the Log Export for slow transfer & confirmation for the USB port slow transfer issue. I kept it in the application repo to keep it safe. Commented out to not ...

This code is going to be used later in the Log Export for slow transfer & confirmation for the USB port slow transfer issue.
I kept it in the application repo to keep it safe.
Commented out to not affect the coverage.

http://dvm-linux02:8060/cru/#UI-DEN-12716-2CFR-30978 in PostTreatmentReview.qml@40

http://dvm-linux02:8060/cru/#UI-DEN-12716-2CFR-30978
in PostTreatmentReview.qml@40

Where would this value change?

Where would this value change?

At the moment I added that switch I assumed it should only be used for V&V and SysV testing.

At the moment I added that switch I assumed it should only be used for V&V and SysV testing.

This is the requirement. PRS 417: The HD Device manufacturing default timezone shall be (UTC-08:00) Pacific Time.

This is the requirement.
PRS 417: The HD Device manufacturing default timezone shall be (UTC-08:00) Pacific Time.

Why is main.cpp all commented out?

Why is main.cpp all commented out?

All systems will be setup for Pacific time zone?

All systems will be setup for Pacific time zone?

Why must this be removed later?

Why must this be removed later?

UI-DEN-12716_UI BN S70
UI-DEN-12716_UI BN S70
DEN-13127: LTS: Development Support

- [Export Button Enabling in VM #61|https://github.com/Diality/testsuites/issues/61]

  1. … 6 more files in changeset.
DEN-13134: IMP: General Implementation for any bugs and any other implementations

DEN-12836: Treatment Report/log export modifications

- Improved the SDItem and renamed to SDCProgressItem.

- Added another separate progress circle component for the USB Driver [USBProgressItem]

- Adding Space check for USB as well to get the USBProgressCircle update.

- Optimized ExportButton and added two size for Settings screen(normal) and Treatment Review screen(small)

- Improved USB button comonent

- Added a global byte size converter in Variable.qml

- Improved MainMenu component

- Added a notification for the TreatmentFlowBase.qml main component.

- Used the notification of the TreatmentFlowBase.qml for the Treatment Review Screen Log export to display feedback to the user.

- Added an individual screen for the log export.

- Removed the Export Log and USB eject from the main settings option screen.

- Added an option for the Export log on the settings home screen which is currently enabled for non-service users too.

- Fixed the DGVersionResponse, log message.

- Put a comment in the FileHandler.cpp about the Qt API QStorageInfo::mountedVolume() which seems so slow on detecting mounted volumes when a USB derive is just plugged in (It could be system issue, since it has also been observed from shell console.)

- Added error handling to the Treatment Log/Report export

DEN-12726: Denali application crashing

- Modified the Logger Maxusage limit to keep 50% of the data/event log and 15% of the service log and keep 35% free, after the log clean up triggered by 15% free storage detected.

DEN-13051: Encounter alarm 6 and SD card detection problem after updating UI version

DEN-13156: Settings screen menu should disappear on service mode

- Improved the MainStack.qml navigation rules.

- Improved the Patient ID keyboard handling, to hide the keyboard, when not on the screen.

- Improved the SettingsStack.qml navigation rules.

DEN-13128: KBM: Cloud Sync Integration

- Saving the CS log in UTC timezone.

DEN-13127: LTS: Development Support

- Fixed the issue found on the End-Treatment screen Confirm button which didn't respect the last instruction page enable only rule.

    • -0
    • +89
    ./SDCProgressItem.qml
    • -0
    • +90
    ./USBProgressItem.qml
  1. … 30 more files in changeset.
DEN-12836: Treatment Report/log export modifications

- DEN-12944: KBM: Usual Meetings/Support/Development

- Modified the setup.sh script to work on PDT (Pacific timezone)

- Added timezone display on the screen by adding it to the VDateTime

- Removed unnecessary FileHandler header include.

- Modified the Treatment Review Export button to export the current TxReport only.

- Added the Safe Eject button for the Post-Treatment TxReport Export.

- Moved Log Folder definitions to the StorageGlobal.

- Fixed the greeting issue [https://github.com/Diality/testsuites/issues/57]

- DEN-11826: Bluetooth BloodPressure Cuff Interval - Observation

- It has completely redesigned

- DEN-9706: Blood Pressure measurement entry window closes automatically

- Auto Hide is disabled

    • -21
    • +20
    ./pages/treatment/TreatmentStack.qml
  1. … 21 more files in changeset.
DEN-12558: HD does not transition to service mode

- implemented the service mode confirmation

- decreased the dialog animation duration

- renamed the service dates MVC to differentiate with service mode.

    • -15
    • +27
    ./pages/settings/SettingsStack.qml
  1. … 30 more files in changeset.
DEN-12541: Get the SD-Card size from the system

  1. … 2 more files in changeset.
DEN-12804: DOX: General review and support

- Removed the compatibility version from the version format

  1. … 3 more files in changeset.
wrapping up multiple sprint codes

- LOG: Added a CLI option to disable the Check-In logs by adding -k, --disable-check-in-log at the CLI. This was interrupting the log tests. The build macro has been removed.

- POST: During the coverage find out that the getter functions have not even been called in the ApplicationPost.

- CloudSync: Added the Treatment Code(TxCode), received from the cloud after the Treatment Report stored by Cloud System. It is a message communication between UI<=(2008)=CS.

- CloudSync: Two error added for the 2008: TxCode

- CloudSync: Due to some issues on Cloud System regarding the credential expiration, It has been asked to add a temporary workaround to send the credentials (1004) before message 1006:Device State, 1007:Treatment Report.

- DEN-12727: Treatment Parameter Screen: Added more space between the parameters components specially the Temperature slider and the option on top.

- DEN-12829: Enabled the back button on the Start Treatment Screen

- DEN-12761: Enabled an empty Patient ID.

It has not yet been implemented on the CloudSync App, so if sent empty CloudSystem will later reject to store the TxReport.

When later implemented an unassigned random Patient ID will be created for that TxReport to be later defined and assigned.

- DEN-12837, DEN-12840: The Disinfection Scheduling Entry and the Confirm button have been removed from the "Dialysate Generator Settings" screen.

- DEN-12837, DEN-12840: The "Dialysate Generator Settings" changed to "DG Cleaning" for the menu option and the screen title.

- Settings: The Clear Alarm Condition button has been removed from the settings screen for the release.

- Settings: The options which has no implementation for the phase 1 have been removed from the Settings screen Service mode options. [Language, Calibration, SW Update, Factory Reset]

- DEN-12841: Reworded the password entries in Service screen and for WiFi Join Screen.

- TxCode, TxReport: removed the commented out Treatment Report Parameters from the code.

- LOG: added one character space int he DEBUG log for the Not Enough Data in Types.

- Settings: The Service Dates default values changed from 0 to "" since they are strings.

- Settings: Added some clarification for the Home screen greetings

  1. … 22 more files in changeset.