VTreatmentCreate.h

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-15911: Post-treatment screen displays heparin type and heparin concentration even if heparin rate and dispensing were disabled in pre-treatment

- Separated the Log and the Gui values for now to address the cloud log later, when the cloud is ready to not to reject Logs to have NONE.

  1. … 8 more files in changeset.
DEN-12842: Updated - UI: prepopulate option on the create treatment screen (Phase 1b)

Remove model refactoring and restored to original

Corrected slider + switch state update for prepopulating

  1. … 5 more files in changeset.
DEN-12842: Added - UI: prepopulate option on the create treatment screen (Phase 1b)

Refactored the adjust parameters model to add a stored prepopulate model

Added resetting and setting selection to gridSelection.qml component

Add a set of properties to store pre-confirmed treatment parameters

Add a model for storing unconfirmed data

Add handling of resetting and pre-populating treatment parameters

  1. … 7 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 54 more files in changeset.
DEN-15359: Usability: Arterial and Venous Pressure Alarm Limits

- DEN-15387: UI: Updates for treatment parameter entry (A/V alarm window)

- DEN-15388: UI: Updates for in-treatment adjustment of alarm parameter window (pressure alarm screen)

- Monitoring components for the TxHome, Rinseback, BloodPrime

  1. … 35 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 9 more files in changeset.
DEN-14955 Tx Create Reject

- code enhancement

- option buttons rejection improvement

  1. … 5 more files in changeset.
DEN-14955 Tx Create Rejection - Include the type buttons' reject reason - Modify to make gridSelection title color changeable

  1. … 4 more files in changeset.
DEN-14955 TxCreateRejection - Update based on feedback from Behrouz - Fix positioning of the invalid parameter after FW response to avoid abnormal repositioning on slider update - Add a constant variable for the no reject reason - Remove modification to VALUESET and add trigger macro call to the VTreatmentCreate file

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 608 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 34 more files in changeset.
DEN-13962: UI: Update pressure limits to match treatment params and UI screen

- Completely restructure the vTreatmentCreate class and removed all the constant ranges properties from that to the VTreatmentRanges.

- Now reading all the constant ranges values from the DataList.conf from the ui.config repository.

- Used the RANGESET property type for the ranges in the VTreatmentRanges class to have by default properties for min,max, res(olution),def(ault) properties.

- Added an ability to slider to hide the handler when needed (Pre-Treatment)

- Hide the Slider handler when slider switch is off

- Added various PROPERTYs, like IDBASED, RAGESET, VALUESET

- Modified the MPreTreatmentAdjustRequest, duration parameter to treatmentDuration for clarifying.

- Modified the Settings class to able to parse a comment on the Group line.

- Updated the ConfirmTreatmentTable to hold the required labels and added translate function for them. removed the call for get names for performance, since those never change.

- Made the SettingsData connection, standard in the VSettings class.

  1. … 17 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 95 more files in changeset.
DEN-13049: UI Treatment Parameters Entry: Some values for treatment parameter entry on treatment parameter entry screen in create treatment do not match SRS/PRS

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 159 more files in changeset.
DEN-11812 UI Keyboard displays text on the spacebar and has a wink emoji button

- added the keyboard layout and style to remove some buttons and remove the text on the spacebar.

- added DEBUG_DISABLE_MAXIMIZE, due to touch screen issue on some devices, the maximized alarm can't be minimized to see the screen underneath.

- While fixing some issues in the Dialin/Simulator, added the reject reason notification to the Treatment review screen and cleared the values on rejection.

- update the Dialysate Temperature max to the 37 from 38.

  1. … 40 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 584 more files in changeset.
DEN-11814: Heparin Minimum Dispensing Parameters Should Be 0.2

- Reverted back the Heparin min to 0.0 due to the slider's issue which disables the OFF feature, since the value 0.0 becomes an invalid.

- A minor update on the POST progress display to start when 0 to max of complete enum -1.

- An idea of fixing the slider touch issue. [Is DISABLED for now, for more test]

  1. … 13 more files in changeset.
DEN-11814: Heparin Minimum Dispensing Parameters Should Be 0.2

- updated the min range regarding the recent PRS-347, PRS-348

- Fixed the minimum precision of the heparin.

  1. … 4 more files in changeset.
DEN-10422: Dialysate Flow Rate maximum rate selectable in Treatment does not meet PRS resolution

  1. … 1 more file in changeset.
DEN-8145: Issues with UI Create Treatment screen

Remainder items (summary):

- 3) Discuss whether to keep the manufacturing default on the slider since the user needs to confirm each parameter: Agreed to set to the minimum value

- 7) Arterial Pressure and Venous Pressure should be consistent with other parameters, the unit should be shown at the end: will be updated to have the units located in the title of the slider.

- Also fixed the title coloring when the slider has not been adjusted yet.

  1. … 4 more files in changeset.
Finishing up the story for Firmware and adding remaining other tasks

DEN-10607: Implementation [Screen]

DEN-10609: Implementation [Treatment Log]

DEN-11320: KeborMed [ CloudSync Development ]

- Added the message handling for the Device status

- Adding Error handling

  1. … 8 more files in changeset.
DEN-10607: Implementing Heparin

Completed

- Create Screen

- Confirm Screen

- MVC

- Log

- Dialin

  1. … 11 more files in changeset.
DEN-10607: Implementation [Screen]

- Got the information below from the settings to be consistent with all the other screens and views:

- acidConcentrateOptions

- bicarbonateConcentrateOptions

- dialyzerTypeOptions

- heparinTypeOptions

- Added the heparin to the list as well so later can be used on the create treatment screen.

  1. … 1 more file in changeset.
DEN-10607: Implementation [Screen]

- DEN-10608: Implementation [Navigation]

- DEN-10609: Implementation [Treatment Log]

  1. … 8 more files in changeset.
PRS 464 Update: Treatment Parameter, Stop Heparin Dispensing Before End of Treatment

- const quint32 heparinStopTime_Max = 470 <- 480

DEN-8705: Pre-Treatment Setup Screen

- Usability tests feedback modifications

- A/V Pressure Limits ranges update regarding PRS changes.

  1. … 11 more files in changeset.
DEN-8918: BtC: Implementation [Screens] & DEN-8919: BtC: Implementation [Navigations]

- The view needs to be implemented in C++.

  1. … 18 more files in changeset.
DEN-8316: TxLog: Implementation [Simulator/Testing]

  1. … 53 more files in changeset.
DEN-8145: Issues with UI Create Treatment screen

- Other than the comments bellow, we talked about the "--" was not correct which has been fixed.

1) The default value should not be shown on the slider that's not been configured yet

- Not sure that is has to be done but regardless, default values were not correct and fixed.

2) Dialyzer Temperature should show degree Celcius, not just C

- Dialyzer Temperature now has degree circle

3) Discuss whether to keep the manufacturing default on the slider since user needs to confirm each parameter

4) Concentrates and Dialyzer types should match with SRS /PRS

- Has been updated regarding the latest updates (BiCarb:PRS351, Dilyzer:PRS352)

5) Parameter value bounds back and forth on the slider during adjustment (display touch sensitivity issue)

6) Suggest to add + or - on the slider to provide ability to adjust precisely

7) Arterial Pressure and Venous Pressure should be consistent with other parameters, unit should be shown at the end

8) Suggest to display Arterial/Venous Pressure Limit as Arterial/Venous Low/High Pressure Limit and the current value as low/high mmHg

9) Suggest to display the current value of each parameter to the right of the slider

  1. … 7 more files in changeset.