dd-firmware

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
LDT-3603 updated comment wording

    • -1
    • +1
    /firmware/App/Modes/FPModes/FPOperationModes.c
LDT-3603 initial draft for FP set operation sub mode

    • -0
    • +88
    /firmware/App/Modes/FPModes/FPOperationModes.c
    • -0
    • +2
    /firmware/App/Modes/FPModes/FPOperationModes.h
    • -0
    • +35
    /firmware/App/Modes/FPModes/ModeGenPermeate.c
    • -1
    • +1
    /firmware/App/Modes/FPModes/ModeGenPermeate.h
LDT-2004 Dry bicart fixing merge issue

LDT-3223-dialysate-flow-rate---dd---mode-tel-dev-test : Added testcases for FlowSensor and Safetyshutdown file

All comments should start with capital letter

All comments should start with capital letter

comment alignment issue

comment alignment issue

it should be mentioned what function is used for calculating the equation.

it should be mentioned what function is used for calculating the equation.

above comment applicable here .

above comment applicable here .

if sensor id is not in the finite state enum value , it return 0 and here it calculate for P9

if sensor id is not in the finite state enum value , it return 0 and here it calculate for P9

parameter is sensorId, not sensor

parameter is sensorId, not sensor

Bamboo Commit: Updated DDCommon.h with build versions from Bamboo.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged LDT-3605

LDT-3605 added code in the blood leak driver to reset its state machine once the hardware configuration has changed

LEAHI-DD-FIRMWARE-LDT-3605_(DD) BLD sensor intermittently does not read for B2.0 HW
LEAHI-DD-FIRMWARE-LDT-3605_(DD) BLD sensor intermittently does not read for B2.0 HW
Add default. What should we do if caller passed invalid sensor ID?

Add default. What should we do if caller passed invalid sensor ID?

LDT-3103 special build changes

    • -4
    • +4
    /firmware/App/Modes/ModeGenDialysate.c
updated newCondModel to be a U32 before check then casted to enum after. Note: Calculate conductivity has a default statement that will catch any value that is outside of the model enum ( and execu...

updated newCondModel to be a U32 before check then casted to enum after. Note: Calculate conductivity has a default statement that will catch any value that is outside of the model enum ( and execute standard conductivity model )

Merge branch 'staging' into LDT-3372-v1-conductivity-fixes

LDT-3372: fixes set conductivity model per code review

LEAHI-DD-FIRMWARE-LDT-3603_Ability to set operation sub mode on IOFP sub system
LEAHI-DD-FIRMWARE-LDT-3603_Ability to set operation sub mode on IOFP sub system
Then need to add a default case in that function

Then need to add a default case in that function

If all enumeration constants are non-negative and fit within the range of an int or unsigned int, the compiler has the flexibility to choose either a signed or unsigned type. In many common impleme...

If all enumeration constants are non-negative and fit within the range of an int or unsigned int, the compiler has the flexibility to choose either a signed or unsigned type. In many common implementations, it defaults to a signed int

Looked better before - now misspelled.

Looked better before - now misspelled.

0.0F

0.0F

indent looks 1 space off.

indent looks 1 space off.