LDT-4001: code review comments.

This needs TD source update and also dialin update, keep informed system team about this change as this would impact DD demo running.

This needs TD source update and also dialin update, keep informed system team about this change as this would impact DD demo running.

Please add HDF Pump speed register and add a monitor in the substitution pump controller to compare against the target speed and if the difference is out of range, trigger an alarm.

Please add HDF Pump speed register and add a monitor in the substitution pump controller to compare against the target speed and if the difference is out of range, trigger an alarm.

Substitution pump

Substitution pump

Use ternary to explicitly assign TRUE or FALSE.

Use ternary to explicitly assign TRUE or FALSE.

Use ternary to explicitly assign TRUE or FALSE.

Use ternary to explicitly assign TRUE or FALSE.

Missing comment.

Missing comment.

update comment. next one as well.

update comment. next one as well.

blank line after the banner and maintain alphabetical order of include files.

blank line after the banner and maintain alphabetical order of include files.

Bamboo Commit: Updated DDCommon.h with build versions from Bamboo.

LDT-3958 resoving merge issues

    • -24
    • +288
    /firmware/App/Controllers/Heaters.c
LDT-4001: previous CR comments. updates to TD interface and messaging

LEAHI-DD-FIRMWARE-LDT-4001_HDF - FW Implementation - 2/4:
LEAHI-DD-FIRMWARE-LDT-4001_HDF - FW Implementation - 2/4:
Done.

Done.

LDT-3958 disabling blood leak alarm

LDT-4001: fixes and merges from previous working branch.

    • -0
    • +73
    /firmware/App/Controllers/SubstitutionPump.h
LDT-4000: changed beta 2.0 test config to beta 1.9 test config

    • -12
    • +12
    /firmware/App/Monitors/Temperature.c
I think we also need this to be uncommented, as we were running into an alarm because of this.

I think we also need this to be uncommented, as we were running into an alarm because of this.

LDT-3958 disabling few alarms

LDT-3958 alarm disabling

Added new field for d1 filtered temperature and restored the pressure sensor temperature.

Added new field for d1 filtered temperature and restored the pressure sensor temperature.

The latest code from staging has an update and fixes this comment. It's not showing here, since the file was already reviewed.

The latest code from staging has an update and fixes this comment. It's not showing here, since the file was already reviewed.

as discussed, these were outcome of the test results. we will review these gains with latest controller.

as discussed, these were outcome of the test results. we will review these gains with latest controller.

Done.

Done.

Done.

Done.

LDT-3223-dialysate-flow-rate---dd---mode-tel-dev-test-3-30-2026 : Added testcases to TaskGeneral

LDT-3223-dialysate-flow-rate---dd---mode-tel-dev-test-3-30-2026 : Added test cases for PersistentAlarm.c file

LDT-3958 heater control update

    • -15
    • +1
    /firmware/App/Controllers/Heaters.c