Controllers

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
LDT-4873 latest from staging

LDT-4873 fixing cond level sensor alarm

Same question as above: should we alarm if the retries failed?

Same question as above: should we alarm if the retries failed?

This else is not needed because isFlashAddressValid is set to False at the top of the function (Optional to change).

This else is not needed because isFlashAddressValid is set to False at the top of the function (Optional to change).

This else is not needed because isFlashAddressValid is set to False at the top of the function (Optional to change).

This else is not needed because isFlashAddressValid is set to False at the top of the function (Optional to change).

If a retry goes wrong will you trigger an alarm?

If a retry goes wrong will you trigger an alarm?

Any fault alarm here?

Any fault alarm here?

Have these variables been initialized correctly?

Have these variables been initialized correctly?

Remove extra blank line.

Remove extra blank line.

We generally don't add { } to switch cases unless they require local variable declarations.

We generally don't add { } to switch cases unless they require local variable declarations.

If we're not including a timestamp, can we just remove this line of code?

If we're not including a timestamp, can we just remove this line of code?

Why commented out?

Why commented out?

merged LDT-4410 Into develop

  1. … 1 more file in changeset.
merged latest staging

  1. … 1 more file in changeset.
0.0F(float type).

0.0F(float type).

I meant 'handleBicarbChamberCartridgeFillWaterEndState function stops the inlet water to cartridge'.

I meant 'handleBicarbChamberCartridgeFillWaterEndState function stops the inlet water to cartridge'.

Stepper driver firmware implementation

  1. … 1 more file in changeset.
merged LDT-4410 to develop

  1. … 1 more file in changeset.
Delete it.

Delete it.

LDT-4410 code review comments addressed

  1. … 1 more file in changeset.
Remove blank spaces after =.

Remove blank spaces after =.

To be deleted this line?

To be deleted this line?

Add a comment here to wait for BC switching to go Idle state and transition to spent chamber fill state. Also, add a space after (). Apply the same comment in the next Dialysate delivery state as w...

Add a comment here to wait for BC switching to go Idle state and transition to spent chamber fill state. Also, add a space after (). Apply the same comment in the next Dialysate delivery state as well.

I guess we can remove the "balancingChamberFillInProgress" check here.

I guess we can remove the "balancingChamberFillInProgress" check here.

Driver update code part 1

  1. … 2 more files in changeset.
Remove the TODO comment.

Remove the TODO comment.

align the comment.

align the comment.

Vendor - TEL - Sameer PoyilWhat's the value you are recommending for?

Vendor - TEL - Sameer PoyilWhat's the value you are recommending for?

Remove values in the comment, as they are changing during development (200ms persistence, 4 PSI low pressure, 17 PSI high pressure etc.,), instead use defined persistence, low or high threshold pre...

Remove values in the comment, as they are changing during development (200ms persistence, 4 PSI low pressure, 17 PSI high pressure etc.,), instead use defined persistence, low or high threshold pressure etc.,