BloodFlow.c

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-4211: Finished HD pumps open loop Dialin command update. Started data infrustructure for treatment parameters mode.

  1. … 15 more files in changeset.
DEN-4212: Tuned blood and dialysate flow sensors. Changed blood and dialysate pump control back to closed loop. Updated FPGA ID.

  1. … 5 more files in changeset.
DIAL-1: added open loop override capability to HD pumps

  1. … 1 more file in changeset.
DEN-4212, DEN-4217: Added accelerometer feature. Added calibration to blood and dialysate flow. Moved msg payload structures to a separate header file.

  1. … 13 more files in changeset.
Handle CAN and UART error/warning notifications. Differentiated HD from DG s/w fault. Updated some older module/function headers to latest format.

  1. … 24 more files in changeset.
Added f/w debug message support. Added debug messages for CAN/UART error interrupts and alarm trigger/clear events. Added FPGA version info to version message. Added handler for DG/UI alarm trigger/clear events. Added debug test code for task/interrupt timing.

  1. … 16 more files in changeset.
Bamboo Commit: Updated the Copyright with the current year and replaced tabs with 4 spaces

  1. … 70 more files in changeset.
Changed UF reference volume update approach from time x rate equation to accumulation of step time x rate. Fixed debug logging of event data when alarm is triggered. Updated flow test build output.

  1. … 4 more files in changeset.
Updated FPGA register map to latest HDD. Separated ramp up and ramp down rates so they can be set differently for blood and dialysate pumps.

  1. … 6 more files in changeset.
Fixed UF change verification logic. Fixed doxygen comments.

  1. … 3 more files in changeset.
Added test code that can read slow page of FPGA. Added test code for flow signal strength and occlusion sensor debug output for h/w testing. Delayed reservoir switch until after DG has flushed lines and started recirculating in order to get a more stable reading on load cells.

  1. … 7 more files in changeset.
Added minimum 1 hr treatment to settings adjustment range checks. Changed blood/dialysate inlet pump flow rate filter approach. Changed blood/dialysate inlet pump control interval to 10 sec.

  1. … 2 more files in changeset.
Updated UT/IT for BloodFlow, Interrupts, FPGA modules.

  1. … 2 more files in changeset.
Updated UT. Addressed code review feedback.

  1. … 3 more files in changeset.
Added blood pump rotor max alarm.

  1. … 2 more files in changeset.
Added pump alarms and checks.

  1. … 9 more files in changeset.
Added new HD pump alarms. Started implementing new HD pump alarms.

  1. … 7 more files in changeset.
Added calculation of pump speed and direction per hall sensor count from FPGA on all 3 pumps.

  1. … 3 more files in changeset.
Added interrupt handler for rotor hall sensor edges. Handled BP rotor hall sensor edges.

  1. … 13 more files in changeset.
De-macroized some Dialin API code for Vectorcast and doxygen. Updated some IT.

  1. … 3 more files in changeset.
Added messaging for blood/dialysate flow rate change and HD to DG dialysate temperature set points.

  1. … 9 more files in changeset.
Perparing for a merge with DEN-1906

  1. … 46 more files in changeset.
Fixes per UT and IT. Updates to UT and IT.

  1. … 7 more files in changeset.
Fixes for UF and treatment time changes. Updated pump factors.

  1. … 9 more files in changeset.
Changed breadboard build switch to be more biased toward the new HD board. Added UF pause resume functions and message handling. Updated internal ADC voltage and conversion factors for new HD board. Commented out RTC POST until RTC on new HD board is configured.

  1. … 12 more files in changeset.
Added HD specific common header. Streamlined includes for Common.h and HDCommon.h.

  1. … 55 more files in changeset.
CAN Test #1.

  1. … 8 more files in changeset.
Updated UT for pump modules to include open loop mode.

  1. … 5 more files in changeset.
Added dialysis sub-mode. Misc fixes.

  1. … 25 more files in changeset.
Added DPo implementation. Added some messagine for DPo. Added FIRFilters. Updated HALCoGen pinouts.

  1. … 38 more files in changeset.