Controllers

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
LDT-2116 addressed rewview comments

  1. … 3 more files in changeset.
Fixed reset during fill and full. Adjusted samples for duty cycle average calculation.

  1. … 1 more file in changeset.
LDT-2116: pushed new closed loop coefficients for tgt flow of 750mL

LDT-2116 changed Gen logic to go open loop for 3 sec and transition to closed loop.

  1. … 1 more file in changeset.
LDT-1773: code review comments.

LDT-2116: code review updates. Change target flow rate from 700 to 750 per system spec. Reverted debugging broadcast for RO Pump.

  1. … 3 more files in changeset.
LDT-2116 added private def for RO pump target flow

LDT-2116: possible fix for open loop to target flow transition.

LDT-2116: updates to ro pump coefficients. updates to permeate tank control while P39 is closed.

  1. … 1 more file in changeset.
LDT-1773: updated comments for fpga interface. changed valve pwm period to correct value

  1. … 2 more files in changeset.
LDT-1773: implemented FPGA interface for valve PWM. implemented in valves module

  1. … 2 more files in changeset.
LDT-1809 Resolved conflicts while merging staging

  1. … 3 more files in changeset.
LDT-1242: code review comments. comment aligning. fix ro pump control calling set pwm pct when it should have been pwm cnt

  1. … 3 more files in changeset.
LDT-1242: reverted RO controller publishing to no longer contain controller debug data

LDT-1242: updated tank full timing

LDT-1242: reverted flow control output for develop builds as pwm output requires tuning

LDT-1242: opening P34 on tank full state.

LDT-1242: converted RO flow control back to outputting PWM percentages instead of target mL/min.

LDT-1242: updated flow control curve. Renamed fp def op modes. Closing P39 on TANK FULL state

  1. … 4 more files in changeset.
Fixed indentation and review comments

fixed review comments

  1. … 2 more files in changeset.
Fixed the issue to set targets to zero when a firmware call is made for set functions per review comments.

  1. … 4 more files in changeset.
LDT-566: updated RO pwm flow curve from linear to logarithmic.

LDT-566: updated RO pwm flow curve from linear to logarithmic.

removed commented code per review comments

declared result at the top of the scope for test functions

LDT-1242: updated permeate tank control to change based on low level.

  1. … 1 more file in changeset.
LDT-566: reverted publishing change for fine tuning.

LDT-566: fixed RO Pump publishing to no longer have debug data for PI control

LDT-2003: 1st stage of flow control. still wip. moved event from bad crc error to all faults.

  1. … 2 more files in changeset.