AlarmDefs.h

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
blank line before return.

blank line before return.

Blank line after declarations.

Blank line after declarations.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'staging' into LDT-1903-air-trap-level-control---td---0

fixed in latest code

fixed in latest code

implemented

implemented

fixed

fixed

LDT-1903 removing ALARM_ID_TD_AIR_PUMP_DUTY_CYCLE_OUT_OF_RANGE

LDT-1903 Adding Air pump duty cycle out of range alarm

I think this would be a TD s/w fault (alarm already define). Just need a new TD s/w fault ID in AlarmMgmtSWFaults.h.

I think this would be a TD s/w fault (alarm already define). Just need a new TD s/w fault ID in AlarmMgmtSWFaults.h.

This structure doesn't seem necessary - it's just a middle man. Why not just pass the #defines directly into the quadratic macro?

This structure doesn't seem necessary - it's just a middle man. Why not just pass the #defines directly into the quadratic macro?

Fixed

Fixed

removed from here

removed from here

fixed

fixed

We need new alarm . That should be included in SRS right ? or shall i use SET_ALARM_WITH_2_U32_DATA( ALARM_ID_TD_SOFTWARE_FAULT) ?

We need new alarm . That should be included in SRS right ? or shall i use SET_ALARM_WITH_2_U32_DATA( ALARM_ID_TD_SOFTWARE_FAULT) ?

Add blank line after declarations.

Add blank line after declarations.

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Add blank line after declarations.

Add blank line after declarations.

I think alarm should be handled inside the setAirPumpState() function.

I think alarm should be handled inside the setAirPumpState() function.

Should we have alarm here?

Should we have alarm here?

duty misspelled.

duty misspelled.

Can we remove inclusion of fpga here? It shouldn't be needed.

Can we remove inclusion of fpga here? It shouldn't be needed.