Should we add a TODO to remove this one as well?

Should we add a TODO to remove this one as well?

If we are removing this call, then where are we calling the function testDDFloaterLevelStateOverride?

If we are removing this call, then where are we calling the function testDDFloaterLevelStateOverride?

Please update the function name : testDDFloaterLevelStateOverride

Please update the function name : testDDFloaterLevelStateOverride

Why are we removing this TODO comment? It looks like we should keep it.

Why are we removing this TODO comment? It looks like we should keep it.

Do we need TODO comments to restore these alarms later?

Do we need TODO comments to restore these alarms later?

What is the point of this message?

What is the point of this message?

Looks like we need a TODO comment to restore these checks later.

Looks like we need a TODO comment to restore these checks later.

LEAHI-DD-FIRMWARE-LDT-4183_Dialysate Composition (Dry Bicarbonate) - FW Implementation - 2/1:
LEAHI-DD-FIRMWARE-LDT-4183_Dialysate Composition (Dry Bicarbonate) - FW Implementation - 2/1:
LEAHI-DD-FIRMWARE-LDT-3978_Switch Chamber F conductive level sensors to broadcast 1 as sensing liquid
LEAHI-DD-FIRMWARE-LDT-3978_Switch Chamber F conductive level sensors to broadcast 1 as sensing liquid
a time out for decaying D66 pressure to 3 PSI after D65 closed

a time out for decaying D66 pressure to 3 PSI after D65 closed

remove space after <= and add space end of function call.

remove space after <= and add space end of function call.

Is this time out after D80 opening or D65 closing state? if it is D65 closed, you need to start the timer after D65 closing.

Is this time out after D80 opening or D65 closing state? if it is D65 closed, you need to start the timer after D65 closing.

if we are not checking level, this can be removed and function name also can be updated accordingly.

if we are not checking level, this can be removed and function name also can be updated accordingly.

Is this a TODO? If so, add TODO to the comment and maybe be a little more descriptive of what needs to be done.

Is this a TODO? If so, add TODO to the comment and maybe be a little more descriptive of what needs to be done.

can we get 'runPIController' inside deadband control check?

can we get 'runPIController' inside deadband control check?

Initialize the 'dryBicartPersistantStartTime' back to zero after timeout done, for the next iterations.

Initialize the 'dryBicartPersistantStartTime' back to zero after timeout done, for the next iterations.

Nope. Since we are here after level is LOW (fill request from balancing chamber control state machine) and not filled yet. So level is still LOW

Nope. Since we are here after level is LOW (fill request from balancing chamber control state machine) and not filled yet. So level is still LOW

First two logics ANDed correct (first cycle and close period). the third one (BC switch only must be ANDed only with close period), not to include first cycle..

First two logics ANDed correct (first cycle and close period). the third one (BC switch only must be ANDed only with close period), not to include first cycle..

Can we check payload.index is equal to D6_LEVL (instead of above comparision)?

Can we check payload.index is equal to D6_LEVL (instead of above comparision)?

Do we need to check if bicarb chamber level is high, ignore opening D80?

Do we need to check if bicarb chamber level is high, ignore opening D80?

This doesn't seem to be correct. Persistence should start once fill complete pressure is achieved, not after D65 opening?

This doesn't seem to be correct. Persistence should start once fill complete pressure is achieved, not after D65 opening?

Align comment

Align comment

Add comment.

Add comment.

LEAHI-DD-FIRMWARE-LDT-3975_Water Treatment and Degassing - FW Iteration - 4/4:
LEAHI-DD-FIRMWARE-LDT-3975_Water Treatment and Degassing - FW Iteration - 4/4:
Remove space after 'Supply'.

Remove space after 'Supply'.