Add unit of measure to definition name (e.g. _PSI). Should be done for all definitions w/ a unit of measure to prevent unit errors.

Add unit of measure to definition name (e.g. _PSI). Should be done for all definitions w/ a unit of measure to prevent unit errors.

LEAHI-DD-FIRMWARE-LDT-3980_Dialysate Composition - FW Implementation (Dry Bicarbonate) - 1/1:
LEAHI-DD-FIRMWARE-LDT-3980_Dialysate Composition - FW Implementation (Dry Bicarbonate) - 1/1:
fixed

fixed

Align the comments.

Align the comments.

LEAHI-DD-FIRMWARE-LDT-3984_(DD, SW) D46 reads 0 with and without liquid in chamber
LEAHI-DD-FIRMWARE-LDT-3984_(DD, SW) D46 reads 0 with and without liquid in chamber
LEAHI-DD-FIRMWARE-LDT-2004_Dialysate Composition - DD
LEAHI-DD-FIRMWARE-LDT-2004_Dialysate Composition - DD
LEAHI-DD-FIRMWARE-LDT-2004_Dialysate Composition - DD
LEAHI-DD-FIRMWARE-LDT-2004_Dialysate Composition - DD
Its static function and not used by external API to pass the wrong id.

Its static function and not used by external API to pass the wrong id.

0.0F is made by Sean.

0.0F is made by Sean.

by the end of S27 we will know if we need to keep this module. If so, then we can integrate it into PIControllers.c

by the end of S27 we will know if we need to keep this module. If so, then we can integrate it into PIControllers.c

If this is the core change, can we introduce this logic in PI controller.c itself with a simple flag check (say, if 'IsDerivativeEnabled" set to true, add the derivative signal to Control signal. e...

If this is the core change, can we introduce this logic in PI controller.c itself with a simple flag check (say, if 'IsDerivativeEnabled" set to true, add the derivative signal to Control signal. else, just a PI control signal would work. (something similar to Feed forward enabled flag)

nope. removed.

nope. removed.

fixed

fixed

fixed

fixed

fixed

fixed

fixed

fixed

Same comment applies here (removing resetPIController).

Same comment applies here (removing resetPIController).

I think we can remove the balancing chamber fill in progress check for Opening D80 valve, as this does not impact the balancing chamber fill process.

I think we can remove the balancing chamber fill in progress check for Opening D80 valve, as this does not impact the balancing chamber fill process.

Maintain space after and before parenthesis across the file.

Maintain space after and before parenthesis across the file.

Do we need to add another else statement and address an alarm if wrong mixId is passed? or Keep only else part instead of else if ( ACID_MIX_ID == mixID) check?? this change may be needed across th...

Do we need to add another else statement and address an alarm if wrong mixId is passed? or Keep only else part instead of else if ( ACID_MIX_ID == mixID) check?? this change may be needed across the file.

Remove this comment.

Remove this comment.