Remove this condition to apply the speed for BC switch only as well.
Check the else portion and do the required changes.
Name something similar to 'BCSwitchingBasedOnClosePeriodCounter'?
I guess, all these checks should be moved to FillEndState.
after one state of BC switching completed. update the next comment as well.
Should we add a TODO to remove this one as well?
If we are removing this call, then where are we calling the function testDDFloaterLevelStateOverride?
Please update the function name : testDDFloaterLevelStateOverride
Why are we removing this TODO comment? It looks like we should keep it.
Do we need TODO comments to restore these alarms later?
What is the point of this message?
Looks like we need a TODO comment to restore these checks later.
a time out for decaying D66 pressure to 3 PSI after D65 closed
remove space after <= and add space end of function call.
Is this time out after D80 opening or D65 closing state? if it is D65 closed, you need to start the timer after D65 closing.
if we are not checking level, this can be removed and function name also can be updated accordingly.
Is this a TODO? If so, add TODO to the comment and maybe be a little more descriptive of what needs to be done.
can we get 'runPIController' inside deadband control check?