For testing purpose, Since UI board had old build so it was throwing the alarms
I see in the commits that you commented this #define was that on purpose?
Per our coding standards, it should be "0 == currentAirPumpRPM".
Yes removed
Agree! Fixed
I don't think this is needed. You can't get to this else until the delay counter reaches this count.
Is this necessary? Treatment paused state will do this.
Fixed
add airTrapLowerEventCount to input
add airTrapLowerEventStartTime and airTrapLowerEventCount to Output
Add to input and output: airTrapLowerEventCount, airTrapLowerEventStartTime
Add alarm : ALARM_ID_TD_REPEATED_AIR_TRAP_LOWER_EVENTS
break it into two lines
remove flag from the comments