DryBiCart.h

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
LEAHI-DD-FIRMWARE-LDT-4246_FW Support for UI Demo
LEAHI-DD-FIRMWARE-LDT-4246_FW Support for UI Demo
Use the get Acid mix control interval?

Use the get Acid mix control interval?

Updates the comments for all new variables.

Updates the comments for all new variables.

There should be a \b before the type of details (Inputs and Outputs here) for documentation please update for all

There should be a \b before the type of details (Inputs and Outputs here) for documentation please update for all

Remove blank line

Remove blank line

Remove blank line There should be a \b before the type of details (Inputs and Outputs here) for documentation

Remove blank line
There should be a \b before the type of details (Inputs and Outputs here) for documentation

There should be a \b before the type of details (Inputs and Outputs here) for documentation Inputs: None?

There should be a \b before the type of details (Inputs and Outputs here) for documentation
Inputs: None?

LDT-4183 addressed self review comments plus dry bicarb implementation

  1. … 1 more file in changeset.
LEAHI-DD-FIRMWARE-LDT-4183_Dialysate Composition (Dry Bicarbonate) - FW Implementation - 2/1:
LEAHI-DD-FIRMWARE-LDT-4183_Dialysate Composition (Dry Bicarbonate) - FW Implementation - 2/1:
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 12 more files in changeset.
a time out for decaying D66 pressure to 3 PSI after D65 closed

a time out for decaying D66 pressure to 3 PSI after D65 closed

remove space after <= and add space end of function call.

remove space after <= and add space end of function call.

Is this time out after D80 opening or D65 closing state? if it is D65 closed, you need to start the timer after D65 closing.

Is this time out after D80 opening or D65 closing state? if it is D65 closed, you need to start the timer after D65 closing.

if we are not checking level, this can be removed and function name also can be updated accordingly.

if we are not checking level, this can be removed and function name also can be updated accordingly.

Is this a TODO? If so, add TODO to the comment and maybe be a little more descriptive of what needs to be done.

Is this a TODO? If so, add TODO to the comment and maybe be a little more descriptive of what needs to be done.

can we get 'runPIController' inside deadband control check?

can we get 'runPIController' inside deadband control check?

LDT-3978 addressing review comments

  1. … 1 more file in changeset.
LDT-3978 fixing compilation error

  1. … 2 more files in changeset.
Initialize the 'dryBicartPersistantStartTime' back to zero after timeout done, for the next iterations.

Initialize the 'dryBicartPersistantStartTime' back to zero after timeout done, for the next iterations.

Nope. Since we are here after level is LOW (fill request from balancing chamber control state machine) and not filled yet. So level is still LOW

Nope. Since we are here after level is LOW (fill request from balancing chamber control state machine) and not filled yet. So level is still LOW

LDT-3978 firmware implementation for ocean spray

  1. … 3 more files in changeset.
Can we check payload.index is equal to D6_LEVL (instead of above comparision)?

Can we check payload.index is equal to D6_LEVL (instead of above comparision)?

Do we need to check if bicarb chamber level is high, ignore opening D80?

Do we need to check if bicarb chamber level is high, ignore opening D80?

This doesn't seem to be correct. Persistence should start once fill complete pressure is achieved, not after D65 opening?

This doesn't seem to be correct. Persistence should start once fill complete pressure is achieved, not after D65 opening?