TubeSetInstall.c

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Should this check be moved to the fluid bolus exec function?

Should this check be moved to the fluid bolus exec function?

where are we calling an alarm? i think we are only sending event change data in completeBolusToCumulative() right?

where are we calling an alarm?
i think we are only sending event change data in completeBolusToCumulative() right?

do we need this block?

do we need this block?

update the comment as required: ml or L?

update the comment as required: ml or L?

rejReason = INVALID REQUEST FORMAT

rejReason = INVALID REQUEST FORMAT

I think these remaining 4 else ifs need to have another && condition that your service is in idle state so that we don't change back into one of these 4 states while a bolus is in progress.

I think these remaining 4 else ifs need to have another && condition that your service is in idle state so that we don't change back into one of these 4 states while a bolus is in progress.

Can we restore the real code here yet? Ask Dara.

Can we restore the real code here yet? Ask Dara.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 14 more files in changeset.
LDT-1886: Update event related code

  1. … 1 more file in changeset.
We can rename priorState to previousInstallState. Also, it should be declared as a static variable to be able to retain previousState value when the control goes out of the function.

We can rename priorState to previousInstallState.
Also, it should be declared as a static variable to be able to retain previousState value when the control goes out of the function.

LDT-1886: Add send event

  1. … 1 more file in changeset.
maybe use target rate for both.

maybe use target rate for both.

Use #define for array size.

Use #define for array size.

Make the 6 a #define

Make the 6 a #define

only 1 return statement

only 1 return statement

remove comma

remove comma

signal functions should return BOOL.

signal functions should return BOOL.

Move up with rest

Move up with rest

Move this function to your service.

Move this function to your service.

Not needed.

Not needed.

Should always exec your bolus sm.

Should always exec your bolus sm.

Keep 2 blank lines before banner

Keep 2 blank lines before banner

Add fluid bolus state

Add fluid bolus state