Move below Messaging.h to keep list in alphabetical order.

Move below Messaging.h to keep list in alphabetical order.

balancingError is an input.

balancingError is an input.

Same here. Let's use only D29.

Same here. Let's use only D29.

Suppose to use D74, Instead of D17 (for Control only).

Suppose to use D74, Instead of D17 (for Control only).

Move these functions to top area where all public functions defined. let's keep the static function after the public function.

Move these functions to top area where all public functions defined. let's keep the static function after the public function.

Missing Comment.

Missing Comment.

Add previousHeaterState to input and output

Add previousHeaterState to input and output

add prevDryBicartState to input and output.

add prevDryBicartState to input and output.

Add prevBalChamberState to input and output

Add prevBalChamberState to input and output

Add isBalChamberSwitchingActive to input

Add isBalChamberSwitchingActive to input

yes

yes

yes

yes

yes

yes

Align it with declration.

Align it with declration.

Add comments.

Add comments.

LEAHI-DD-FIRMWARE-LDT-4719_Dialysate Composition (Mixing) - FW Implementation - 4/1:
LEAHI-DD-FIRMWARE-LDT-4719_Dialysate Composition (Mixing) - FW Implementation - 4/1:
LEAHI-DD-FIRMWARE-LDT-4873_(Request) Add config to disable conductive level sensor fault
LEAHI-DD-FIRMWARE-LDT-4873_(Request) Add config to disable conductive level sensor fault
LEAHI-DD-FIRMWARE-LDT-3902_Allow DD balancing chamber switch only command to go down to 50 mL/min
LEAHI-DD-FIRMWARE-LDT-3902_Allow DD balancing chamber switch only command to go down to 50 mL/min
Same as above.

Same as above.

Add an event here, when time out happens. Event data should capture fresh side and spent side fill completion status.

Add an event here, when time out happens. Event data should capture fresh side and spent side fill completion status.

You may need to comment/ remove this condition. the pressure stabilized flag should set only when pressure is met (not on chamber fill request)

You may need to comment/ remove this condition. the pressure stabilized flag should set only when pressure is met (not on chamber fill request)

LEAHI-DD-FIRMWARE-LDT-4612_Dialysate Composition (Dry Bicarbonate) - FW Iteration - 4/1:
LEAHI-DD-FIRMWARE-LDT-4612_Dialysate Composition (Dry Bicarbonate) - FW Iteration - 4/1:
it shouldn't be the default value

it shouldn't be the default value

LEAHI-DD-FIRMWARE-LDT-4734_Water Treatment and Degassing - FW Iteration - 5/4:
LEAHI-DD-FIRMWARE-LDT-4734_Water Treatment and Degassing - FW Iteration - 5/4:
LEAHI-DD-FIRMWARE-LDT-4733_(Request) Add variables to capture AC/DC heater electrical data
LEAHI-DD-FIRMWARE-LDT-4733_(Request) Add variables to capture AC/DC heater electrical data