Return type F32.

Return type F32.

Alarm with F32.

Alarm with F32.

U08 to F32?

U08 to F32?

Power is F32?

Power is F32?

I guess power is F32, so size of ( F32 )?

I guess power is F32, so size of ( F32 )?

Just curious, during stall condition RPM completely turns down to zero or close to zero (positive integer)?

Just curious, during stall condition RPM completely turns down to zero or close to zero (positive integer)?

Use ternary operator to assign true or false.

Use ternary operator to assign true or false.

No, Fixed

No, Fixed

I don't think you should be subtracting venMinOffset anymore. Check with Eliza.

I don't think you should be subtracting venMinOffset anymore. Check with Eliza.

LEAHI-TD-FIRMWARE-LDT-4094_Pressures Monitoring - FW Iteration - 1/1:
LEAHI-TD-FIRMWARE-LDT-4094_Pressures Monitoring - FW Iteration - 1/1:
Just use 0x0 and 0xFFFFFFFF

Just use 0x0 and 0xFFFFFFFF

NUM_OF_BATTERY_REGISTERS - 1

NUM_OF_BATTERY_REGISTERS - 1

Put 2 items at end of the list here: BATTERY_END_OF_LIST_TIME_TO_PUBLISH NUM_OF_BATTERY_REGISTERS

Put 2 items at end of the list here:
BATTERY_END_OF_LIST_TIME_TO_PUBLISH
NUM_OF_BATTERY_REGISTERS

Put this one back to NUM_OF_BATTERY_DEVICES.

Put this one back to NUM_OF_BATTERY_DEVICES.

Don't need min/max for this. Just use 0x0 and 0xFFFFFFFF where needed.

Don't need min/max for this. Just use 0x0 and 0xFFFFFFFF where needed.

I don't think we need these min/max #defines. Just use 0x0 for minimums and 0xFFFFFFFF for maximums.

I don't think we need these min/max #defines. Just use 0x0 for minimums and 0xFFFFFFFF for maximums.

Should this check be moved to the fluid bolus exec function?

Should this check be moved to the fluid bolus exec function?

LEAHI-TD-FIRMWARE-LDT-3215_Measured venous pressure decreases to atmosphere if H12 air pump is dead headed
LEAHI-TD-FIRMWARE-LDT-3215_Measured venous pressure decreases to atmosphere if H12 air pump is dead headed
where are we calling an alarm? i think we are only sending event change data in completeBolusToCumulative() right?

where are we calling an alarm?
i think we are only sending event change data in completeBolusToCumulative() right?

do we need this block?

do we need this block?

update the comment as required: ml or L?

update the comment as required: ml or L?