LEAHI-DD-FIRMWARE-LDT-3103

Delete the macro if not being used.

Delete the macro if not being used.

removed

removed

updated comment.

updated comment.

removed.

removed.

removed and added comments.

removed and added comments.

changed.

changed.

done. moved them to dosing concentrates state.

done. moved them to dosing concentrates state.

Remove this reference (across the file) if the implementation is updated.

Remove this reference (across the file) if the implementation is updated.

Update the deadband timing in the comments.

Update the deadband timing in the comments.

Remove blank line. Also, add comments on the key logics for reference.

Remove blank line. Also, add comments on the key logics for reference.

Default case rise count should be 2, instead of 1?

Default case rise count should be 2, instead of 1?

Now that we go with pressure slope detection logic, we don't want to use pressure-based detection for Low Qd. correct?

Now that we go with pressure slope detection logic, we don't want to use pressure-based detection for Low Qd. correct?

The BC valves not even opened and collecting the spent dial pressure as last previous spent dial pressure may not be correct.

The BC valves not even opened and collecting the spent dial pressure as last previous spent dial pressure may not be correct.

updated.

updated.

updated.

updated.

More implied boolean assignments.

More implied boolean assignments.

Implied boolean assignment. Use ternary operator to assign TRUE or FALSE explicitly or don't initialize in declaration and use if else below to assign.

Implied boolean assignment. Use ternary operator to assign TRUE or FALSE explicitly or don't initialize in declaration and use if else below to assign.

done

done

done

done

updated

updated

the else part here, should handle greater than Qd: 300?

the else part here, should handle greater than Qd: 300?

Get this logic into another function and call the function here.

Get this logic into another function and call the function here.

Is this supposed to be difference in pressure, not the spent dial pressure?

Is this supposed to be difference in pressure, not the spent dial pressure?

LEAHI-DD-FIRMWARE-LDT-3103_implement closed loop control for d48 pump speed
LEAHI-DD-FIRMWARE-LDT-3103_implement closed loop control for d48 pump speed
LEAHI-DD-FIRMWARE-LDT-3103_implement closed loop control for d48 pump speed
LEAHI-DD-FIRMWARE-LDT-3103_implement closed loop control for d48 pump speed