pretreatment

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-14209 - UI: Dry self test modifications

- Updated the View and the qml code regarding the changes to the enum made in common HDDefs.h

  1. … 5 more files in changeset.
DEN-13962: UI: Update pressure limits to match treatment params and UI screen

- removed the QML duplicate definitions of arterial/venous

- added monitoring values and used them in monitoring components of arterial/venous.

  1. … 17 more files in changeset.
DEN-13962: UI: Update pressure limits to match treatment params and UI screen

- Completely restructure the vTreatmentCreate class and removed all the constant ranges properties from that to the VTreatmentRanges.

- Now reading all the constant ranges values from the DataList.conf from the ui.config repository.

- Used the RANGESET property type for the ranges in the VTreatmentRanges class to have by default properties for min,max, res(olution),def(ault) properties.

- Added an ability to slider to hide the handler when needed (Pre-Treatment)

- Hide the Slider handler when slider switch is off

- Added various PROPERTYs, like IDBASED, RAGESET, VALUESET

- Modified the MPreTreatmentAdjustRequest, duration parameter to treatmentDuration for clarifying.

- Modified the Settings class to able to parse a comment on the Group line.

- Updated the ConfirmTreatmentTable to hold the required labels and added translate function for them. removed the call for get names for performance, since those never change.

- Made the SettingsData connection, standard in the VSettings class.

    • -201
    • +75
    ./create/PreTreatmentCreate.qml
  1. … 17 more files in changeset.
DEN-13939: IMP: Fixed the merge issue

- Not sure why git didn't get the changes on the ModalDialog's Norification changes to NotificationSmall, and caused the issue on the property IconVisible.

  1. … 8 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 94 more files in changeset.
Merge branch 'DEN-13198-UI_BN_75' into staging

  1. … 44 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    ./connection/PreTreatmentConnectionStack.qml
  1. … 96 more files in changeset.
DEN-13845: CRV: Merge issue of DEN-12716-UI_BN_S70 into staging

- Due to some enum changes, UI couldn't build

- An idea implemented to detect the enum changes by compiler. The implementation may be able to improve.

  1. … 12 more files in changeset.
Fixed the issue on the PatientID confirm text

  1. … 1 more file in changeset.
DEN-12838: UI shall store the latest alarm volume and resend that to HD on POST

- The save ability has been added to the Settings.

- Note: the design of the settings need to change: 1- singleton class should be the controller not the mode, 2 - The column based keyvalue has to change to tree(map).

  1. … 12 more files in changeset.
DEN-13049: UI Treatment Parameters Entry: Some values for treatment parameter entry on treatment parameter entry screen in create treatment do not match SRS/PRS

- A bug found on multiple, ON,OFF toggle, slider sends minimum instead of '0' while ON.

- Slider improved.

- Added float rounding of the request message logs to 3.

  1. … 6 more files in changeset.
DEN-13049 - UI Treatment Parameters Entry: Some values for treatment parameter entry on treatment parameter entry screen in create treatment do not match SRS/PRS

- There was a bug in the fix which was related to the validation and caused to fail the treatment parameters validation.

- also removed the disable if the SD is not detected correctly. (QStorageInfo is not sending DeviceReady correctly)

- enabled the float rounding to 3 in the log

- Enabled the Create Treatment button rejection notification, since it seems it is not being handled by alarms as presumed.

- Fixed the BloodPressure interval value to display OFF when set to zero.

  1. … 4 more files in changeset.
DEN-9129: Unable to proceed past "Create a Custom Treatment" when Arterial and Venous Pressure (High) Limits maximum values are initially selected

Also modified the SliderCreateTreatment.qml to show by the default the zero value not the OFF since it is only two Heparin Bolus/Dispensing which display OFF for zero value.

  1. … 3 more files in changeset.
DEN-13049: UI Treatment Parameters Entry: Some values for treatment parameter entry on treatment parameter entry screen in create treatment do not match SRS/PRS

  1. … 5 more files in changeset.
DEN-13134: IMP: General Implementation for any bugs and any other implementations

DEN-12836: Treatment Report/log export modifications

- Improved the SDItem and renamed to SDCProgressItem.

- Added another separate progress circle component for the USB Driver [USBProgressItem]

- Adding Space check for USB as well to get the USBProgressCircle update.

- Optimized ExportButton and added two size for Settings screen(normal) and Treatment Review screen(small)

- Improved USB button comonent

- Added a global byte size converter in Variable.qml

- Improved MainMenu component

- Added a notification for the TreatmentFlowBase.qml main component.

- Used the notification of the TreatmentFlowBase.qml for the Treatment Review Screen Log export to display feedback to the user.

- Added an individual screen for the log export.

- Removed the Export Log and USB eject from the main settings option screen.

- Added an option for the Export log on the settings home screen which is currently enabled for non-service users too.

- Fixed the DGVersionResponse, log message.

- Put a comment in the FileHandler.cpp about the Qt API QStorageInfo::mountedVolume() which seems so slow on detecting mounted volumes when a USB derive is just plugged in (It could be system issue, since it has also been observed from shell console.)

- Added error handling to the Treatment Log/Report export

DEN-12726: Denali application crashing

- Modified the Logger Maxusage limit to keep 50% of the data/event log and 15% of the service log and keep 35% free, after the log clean up triggered by 15% free storage detected.

DEN-13051: Encounter alarm 6 and SD card detection problem after updating UI version

DEN-13156: Settings screen menu should disappear on service mode

- Improved the MainStack.qml navigation rules.

- Improved the Patient ID keyboard handling, to hide the keyboard, when not on the screen.

- Improved the SettingsStack.qml navigation rules.

DEN-13128: KBM: Cloud Sync Integration

- Saving the CS log in UTC timezone.

DEN-13127: LTS: Development Support

- Fixed the issue found on the End-Treatment screen Confirm button which didn't respect the last instruction page enable only rule.

  1. … 44 more files in changeset.
DEN-12836: Treatment Report/log export modifications

- DEN-12944: KBM: Usual Meetings/Support/Development

- Modified the setup.sh script to work on PDT (Pacific timezone)

- Added timezone display on the screen by adding it to the VDateTime

- Removed unnecessary FileHandler header include.

- Modified the Treatment Review Export button to export the current TxReport only.

- Added the Safe Eject button for the Post-Treatment TxReport Export.

- Moved Log Folder definitions to the StorageGlobal.

- Fixed the greeting issue [https://github.com/Diality/testsuites/issues/57]

- DEN-11826: Bluetooth BloodPressure Cuff Interval - Observation

- It has completely redesigned

- DEN-9706: Blood Pressure measurement entry window closes automatically

- Auto Hide is disabled

    • -4
    • +11
    ./connection/PreTreatmentConnectionStack.qml
  1. … 26 more files in changeset.
wrapping up multiple sprint codes

- LOG: Added a CLI option to disable the Check-In logs by adding -k, --disable-check-in-log at the CLI. This was interrupting the log tests. The build macro has been removed.

- POST: During the coverage find out that the getter functions have not even been called in the ApplicationPost.

- CloudSync: Added the Treatment Code(TxCode), received from the cloud after the Treatment Report stored by Cloud System. It is a message communication between UI<=(2008)=CS.

- CloudSync: Two error added for the 2008: TxCode

- CloudSync: Due to some issues on Cloud System regarding the credential expiration, It has been asked to add a temporary workaround to send the credentials (1004) before message 1006:Device State, 1007:Treatment Report.

- DEN-12727: Treatment Parameter Screen: Added more space between the parameters components specially the Temperature slider and the option on top.

- DEN-12829: Enabled the back button on the Start Treatment Screen

- DEN-12761: Enabled an empty Patient ID.

It has not yet been implemented on the CloudSync App, so if sent empty CloudSystem will later reject to store the TxReport.

When later implemented an unassigned random Patient ID will be created for that TxReport to be later defined and assigned.

- DEN-12837, DEN-12840: The Disinfection Scheduling Entry and the Confirm button have been removed from the "Dialysate Generator Settings" screen.

- DEN-12837, DEN-12840: The "Dialysate Generator Settings" changed to "DG Cleaning" for the menu option and the screen title.

- Settings: The Clear Alarm Condition button has been removed from the settings screen for the release.

- Settings: The options which has no implementation for the phase 1 have been removed from the Settings screen Service mode options. [Language, Calibration, SW Update, Factory Reset]

- DEN-12841: Reworded the password entries in Service screen and for WiFi Join Screen.

- TxCode, TxReport: removed the commented out Treatment Report Parameters from the code.

- LOG: added one character space int he DEBUG log for the Not Enough Data in Types.

- Settings: The Service Dates default values changed from 0 to "" since they are strings.

- Settings: Added some clarification for the Home screen greetings

    • -0
    • +1
    ./connection/PreTreatmentConnectionStack.qml
  1. … 29 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 159 more files in changeset.
DEN-12512: INT: Support/Meetings/Planning

DEN-09712 Date format between system date/time and BP/HP "Last Read" is not consistent

DEN-09929 Treatment Log shows "?C" instead of "°C"

PRS-35: Regarding Sean's point about the create treatment parameters, made the parameters dynamic by the Configurations/DataList.cfg in ui.config repository.

  1. … 14 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 74 more files in changeset.
DEN-12180: Internal general support

- investigate on the Begin Prime back to Cartridge installation

- Added objectName for easier debug.

- added the stack change for the Cartridge Installation in case FW wants to navigate back from Begin Prime to the beginning of Cartridge Installation.

  1. … 3 more files in changeset.
DEN-10974: Syringe pump installation screen allows user to proceed before syringe installed

- Added the response message 0xAD00(173), for the request message of 0x6200(98):Pre-treatmetn disposables confirm.

  1. … 11 more files in changeset.
DEN-12106: Heparin Delivery Active Text Does Not Appear

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -3
    • +3
    ./connection/PreTreatmentConnectionStack.qml
    • -5
    • +5
    ./connection/PreTreatmentUltrafiltration.qml
  1. … 574 more files in changeset.
UI-BUG-S62: UI-BUG-S62-1

Addressed code reviews.

  1. … 5 more files in changeset.
DEN-11814: Heparin Minimum Dispensing Parameters Should Be 0.2

- Reverted back the Heparin min to 0.0 due to the slider's issue which disables the OFF feature, since the value 0.0 becomes an invalid.

- A minor update on the POST progress display to start when 0 to max of complete enum -1.

- An idea of fixing the slider touch issue. [Is DISABLED for now, for more test]

  1. … 13 more files in changeset.
Merge branch 'UI_BUG_S62' into develop

    • -0
    • +1
    ./connection/PreTreatmentConnectionStack.qml
DEN-11825: BP/HR screen has no instruction for user to read from bluetooth cuff or enter information

- Center aligned the notification

    • -0
    • +1
    ./connection/PreTreatmentConnectionStack.qml
  1. … 1 more file in changeset.
DEN-11814: Heparin Minimum Dispensing Parameters Should Be 0.2

- updated the min range regarding the recent PRS-347, PRS-348

- Fixed the minimum precision of the heparin.

  1. … 4 more files in changeset.
Merge branch 'UI_BUG_S62' into develop

    • -0
    • +9
    ./connection/PreTreatmentConnectionStack.qml