Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 612 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 19 more files in changeset.
CS-5: UI: TxCode generation on device

  1. … 9 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 20 more files in changeset.
DEN-15911: Post-treatment screen displays heparin type and heparin concentration even if heparin rate and dispensing were disabled in pre-treatment

- Separated the Log and the Gui values for now to address the cloud log later, when the cloud is ready to not to reject Logs to have NONE.

  1. … 8 more files in changeset.
DEN-12842: Updated - UI: prepopulate option on the create treatment screen (Phase 1b)

Remove model refactoring and restored to original

Corrected slider + switch state update for prepopulating

  1. … 5 more files in changeset.
DEN-12842: Added - UI: prepopulate option on the create treatment screen (Phase 1b)

Refactored the adjust parameters model to add a stored prepopulate model

Added resetting and setting selection to gridSelection.qml component

Add a set of properties to store pre-confirmed treatment parameters

Add a model for storing unconfirmed data

Add handling of resetting and pre-populating treatment parameters

  1. … 7 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 54 more files in changeset.
DEN-15359: Usability: Arterial and Venous Pressure Alarm Limits

- DEN-15387: UI: Updates for treatment parameter entry (A/V alarm window)

- DEN-15388: UI: Updates for in-treatment adjustment of alarm parameter window (pressure alarm screen)

- Monitoring components for the TxHome, Rinseback, BloodPrime

  1. … 35 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 9 more files in changeset.
DEN-14955 Tx Create Reject

- code enhancement

- option buttons rejection improvement

  1. … 5 more files in changeset.
DEN-14955 Tx Create Rejection - Include the type buttons' reject reason - Modify to make gridSelection title color changeable

  1. … 4 more files in changeset.
DEN-14955 TxCreateRejection - Update based on feedback from Behrouz - Fix positioning of the invalid parameter after FW response to avoid abnormal repositioning on slider update - Add a constant variable for the no reject reason - Remove modification to VALUESET and add trigger macro call to the VTreatmentCreate file

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 608 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 34 more files in changeset.
DEN-13962: UI: Update pressure limits to match treatment params and UI screen

- Completely restructure the vTreatmentCreate class and removed all the constant ranges properties from that to the VTreatmentRanges.

- Now reading all the constant ranges values from the DataList.conf from the ui.config repository.

- Used the RANGESET property type for the ranges in the VTreatmentRanges class to have by default properties for min,max, res(olution),def(ault) properties.

- Added an ability to slider to hide the handler when needed (Pre-Treatment)

- Hide the Slider handler when slider switch is off

- Added various PROPERTYs, like IDBASED, RAGESET, VALUESET

- Modified the MPreTreatmentAdjustRequest, duration parameter to treatmentDuration for clarifying.

- Modified the Settings class to able to parse a comment on the Group line.

- Updated the ConfirmTreatmentTable to hold the required labels and added translate function for them. removed the call for get names for performance, since those never change.

- Made the SettingsData connection, standard in the VSettings class.

  1. … 17 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 95 more files in changeset.
DEN-13049: UI Treatment Parameters Entry: Some values for treatment parameter entry on treatment parameter entry screen in create treatment do not match SRS/PRS

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 159 more files in changeset.
DEN-11812 UI Keyboard displays text on the spacebar and has a wink emoji button

- added the keyboard layout and style to remove some buttons and remove the text on the spacebar.

- added DEBUG_DISABLE_MAXIMIZE, due to touch screen issue on some devices, the maximized alarm can't be minimized to see the screen underneath.

- While fixing some issues in the Dialin/Simulator, added the reject reason notification to the Treatment review screen and cleared the values on rejection.

- update the Dialysate Temperature max to the 37 from 38.

  1. … 40 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 584 more files in changeset.
DEN-11814: Heparin Minimum Dispensing Parameters Should Be 0.2

- Reverted back the Heparin min to 0.0 due to the slider's issue which disables the OFF feature, since the value 0.0 becomes an invalid.

- A minor update on the POST progress display to start when 0 to max of complete enum -1.

- An idea of fixing the slider touch issue. [Is DISABLED for now, for more test]

  1. … 13 more files in changeset.
DEN-11814: Heparin Minimum Dispensing Parameters Should Be 0.2

- updated the min range regarding the recent PRS-347, PRS-348

- Fixed the minimum precision of the heparin.

  1. … 4 more files in changeset.
DEN-10422: Dialysate Flow Rate maximum rate selectable in Treatment does not meet PRS resolution

  1. … 1 more file in changeset.
DEN-8145: Issues with UI Create Treatment screen

Remainder items (summary):

- 3) Discuss whether to keep the manufacturing default on the slider since the user needs to confirm each parameter: Agreed to set to the minimum value

- 7) Arterial Pressure and Venous Pressure should be consistent with other parameters, the unit should be shown at the end: will be updated to have the units located in the title of the slider.

- Also fixed the title coloring when the slider has not been adjusted yet.

  1. … 4 more files in changeset.
Finishing up the story for Firmware and adding remaining other tasks

DEN-10607: Implementation [Screen]

DEN-10609: Implementation [Treatment Log]

DEN-11320: KeborMed [ CloudSync Development ]

- Added the message handling for the Device status

- Adding Error handling

  1. … 8 more files in changeset.
DEN-10607: Implementing Heparin

Completed

- Create Screen

- Confirm Screen

- MVC

- Log

- Dialin

  1. … 11 more files in changeset.
DEN-10607: Implementation [Screen]

- Got the information below from the settings to be consistent with all the other screens and views:

- acidConcentrateOptions

- bicarbonateConcentrateOptions

- dialyzerTypeOptions

- heparinTypeOptions

- Added the heparin to the list as well so later can be used on the create treatment screen.

  1. … 1 more file in changeset.
DEN-10607: Implementation [Screen]

- DEN-10608: Implementation [Navigation]

- DEN-10609: Implementation [Treatment Log]

  1. … 8 more files in changeset.
PRS 464 Update: Treatment Parameter, Stop Heparin Dispensing Before End of Treatment

- const quint32 heparinStopTime_Max = 470 <- 480