Controllers

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
LDT-3978-addressed review comments

LDT-3963 enabled fault mode tranisiton for switching off hetares for dilaysate flow rate testing

  1. … 8 more files in changeset.
remove space after <= and add space end of function call.

remove space after <= and add space end of function call.

Is this time out after D80 opening or D65 closing state? if it is D65 closed, you need to start the timer after D65 closing.

Is this time out after D80 opening or D65 closing state? if it is D65 closed, you need to start the timer after D65 closing.

if we are not checking level, this can be removed and function name also can be updated accordingly.

if we are not checking level, this can be removed and function name also can be updated accordingly.

Is this a TODO? If so, add TODO to the comment and maybe be a little more descriptive of what needs to be done.

Is this a TODO? If so, add TODO to the comment and maybe be a little more descriptive of what needs to be done.

can we get 'runPIController' inside deadband control check?

can we get 'runPIController' inside deadband control check?

LDT-3978 fixing compilation warning

LDT-3978 addressing review comments

LDT-3978 fixing compilation error

  1. … 1 more file in changeset.
LDT-3978 fix for compilation error

Merge branch LDT-3980-dialysate-composition-dry-bicarb---fw-impl into develop

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch LDT-3980-dialysate-composition-dry-bicarb---fw-impl into staging

LDT-3980 addressed review comments

Initialize the 'dryBicartPersistantStartTime' back to zero after timeout done, for the next iterations.

Initialize the 'dryBicartPersistantStartTime' back to zero after timeout done, for the next iterations.

LDT-3980 addressed review comments

Nope. Since we are here after level is LOW (fill request from balancing chamber control state machine) and not filled yet. So level is still LOW

Nope. Since we are here after level is LOW (fill request from balancing chamber control state machine) and not filled yet. So level is still LOW

LDT-3978 firmware implementation for ocean spray

  1. … 1 more file in changeset.
First two logics ANDed correct (first cycle and close period). the third one (BC switch only must be ANDed only with close period), not to include first cycle..

First two logics ANDed correct (first cycle and close period). the third one (BC switch only must be ANDed only with close period), not to include first cycle..

LDT-3975 WIP initial draft for boost pump refactor

  1. … 7 more files in changeset.
Can we check payload.index is equal to D6_LEVL (instead of above comparision)?

Can we check payload.index is equal to D6_LEVL (instead of above comparision)?

Do we need to check if bicarb chamber level is high, ignore opening D80?

Do we need to check if bicarb chamber level is high, ignore opening D80?

This doesn't seem to be correct. Persistence should start once fill complete pressure is achieved, not after D65 opening?

This doesn't seem to be correct. Persistence should start once fill complete pressure is achieved, not after D65 opening?

Align comment

Align comment

Add comment.

Add comment.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

LEAHI-DD-FIRMWARE-LDT-3975_Water Treatment and Degassing - FW Iteration - 4/4:
LEAHI-DD-FIRMWARE-LDT-3975_Water Treatment and Degassing - FW Iteration - 4/4: