LDT-4002: Initial version of Switches module for HDF

  1. … 4 more files in changeset.
LEAHI-DD-FIRMWARE-LDT-4011_Isolated UF (DD) - FW Implementation - 1/2:
LEAHI-DD-FIRMWARE-LDT-4011_Isolated UF (DD) - FW Implementation - 1/2:
LDT-4001: code review comments.

  1. … 1 more file in changeset.
Update alarms and Input/Output sections.

Update alarms and Input/Output sections.

Also, add option individual parameter override for substitutions pump rate as well.

Also, add option individual parameter override for substitutions pump rate as well.

Space after parenthesis. Please add the same in else portion as well (when treatment is in progress, update the rate).

Space after parenthesis. Please add the same in else portion as well (when treatment is in progress, update the rate).

fixed.

fixed.

fixed

fixed

fixed

fixed

fixed

fixed

added

added

LDT-4001: D92 monitoring and alarms with Hall sensor count now in fpga

  1. … 3 more files in changeset.
This needs TD source update and also dialin update, keep informed system team about this change as this would impact DD demo running.

This needs TD source update and also dialin update, keep informed system team about this change as this would impact DD demo running.

Please add HDF Pump speed register and add a monitor in the substitution pump controller to compare against the target speed and if the difference is out of range, trigger an alarm.

Please add HDF Pump speed register and add a monitor in the substitution pump controller to compare against the target speed and if the difference is out of range, trigger an alarm.

Substitution pump

Substitution pump

Missing comment.

Missing comment.

update comment. next one as well.

update comment. next one as well.

blank line after the banner and maintain alphabetical order of include files.

blank line after the banner and maintain alphabetical order of include files.

LDT-4001: previous CR comments. updates to TD interface and messaging

  1. … 2 more files in changeset.
LEAHI-DD-FIRMWARE-LDT-4001_HDF - FW Implementation - 2/4:
LEAHI-DD-FIRMWARE-LDT-4001_HDF - FW Implementation - 2/4:
LDT-4001: fixes and merges from previous working branch.

  1. … 3 more files in changeset.
LDT-4000: changed beta 2.0 test config to beta 1.9 test config

  1. … 5 more files in changeset.
LEAHI-DD-FIRMWARE-LDT-3975_Water Treatment and Degassing - FW Iteration - 4/4:
LEAHI-DD-FIRMWARE-LDT-3975_Water Treatment and Degassing - FW Iteration - 4/4:
LDT-3975: Added distinct s/w fault IDs for PID controller faults.

Merged staging and resolved conflicts

  1. … 10 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.
On DD f/w side, this should just be a simple pump controller. States should be Off and On (control to commanded Qsub). There is no flow sensor for feedback control, so assuming this is open loop si...

On DD f/w side, this should just be a simple pump controller. States should be Off and On (control to commanded Qsub). There is no flow sensor for feedback control, so assuming this is open loop similar to UF pump.
GenDialysate mode will have an HDF sub-state that signals this controller from Off to On and would also call separate controller for heater and separate (Switches) monitor for door sensors.

With all of that in mind, I think this unit should have a simpler name like SubPump.