LDT-3103 addressed code review comments

    • -2
    • +31
    ./App/Controllers/BalancingChamber.c
in the checkSpentFillComplete function we look at the current and the previous pressure sample and calculate the difference to get the adjusted speed for d48 pump.

in the checkSpentFillComplete function we look at the current and the previous pressure sample and calculate the difference to get the adjusted speed for d48 pump.

I am currently using all of them.

I am currently using all of them.

Set D12 pump to 1250 RPM for beta 1.9 and 2.0 during dialysate generation mode.

Done.

Done.

Done.

Done.

Heater control update

LDT-2004: fix for 1.9 changes

LEAHI-DD-FIRMWARE-LDT-3630_Update FW to set D12 Speed to 1250 RPM during dialysate generation mode for B1.9/2.0
LEAHI-DD-FIRMWARE-LDT-3630_Update FW to set D12 Speed to 1250 RPM during dialysate generation mode for B1.9/2.0
Add comments

Add comments

Align indent

Align indent

Align "="

Align "="

LDT-2004: reverted 2.0 fpga init changes.

LDT-2004: override fixes for v1 and v2 sensors.

    • -135
    • +12
    ./App/Drivers/ConductivitySensors.c
FALSE?

FALSE?

if fresh and spent pressure is not met after certain time, we should trigger an alarm and not to stuck in this state forever.

if fresh and spent pressure is not met after certain time, we should trigger an alarm and not to stuck in this state forever.

are we using this variable, if not, can we remove across the file?

are we using this variable, if not, can we remove across the file?

Add comment

Add comment

Remove the unused macros.

Remove the unused macros.

fixed.

fixed.

added.

added.

deleted

deleted

done.

done.

done.

done.

changed and updated all the other places.

changed and updated all the other places.

changed.

changed.

handled accordingly.

handled accordingly.

done

done

done

done