leahi-common

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Maintain alphabetical order.

Maintain alphabetical order.

Eliza says we do allow multiple iso UF sessions in a treatment. So do we still need this else if, but looking at <= iso UF elapsed session time?

Eliza says we do allow multiple iso UF sessions in a treatment. So do we still need this else if, but looking at <= iso UF elapsed session time?

What about HDF?

What about HDF?

Once UI software sends the volume and/or duration change request and once we change them, the UI software will send another request to confirm the displayed settings to the user and that is what is...

Once UI software sends the volume and/or duration change request and once we change them, the UI software will send another request to confirm the displayed settings to the user and that is what is being handled in this function.

return result;

return result;

Keep local result and set to TRUE or FALSE with a ternary.

Keep local result and set to TRUE or FALSE with a ternary.

Which setting does this message confirm (duration or volume)?

Which setting does this message confirm (duration or volume)?

Update alarms and Input/Output sections.

Update alarms and Input/Output sections.

Also, add option individual parameter override for substitutions pump rate as well.

Also, add option individual parameter override for substitutions pump rate as well.

Space after parenthesis. Please add the same in else portion as well (when treatment is in progress, update the rate).

Space after parenthesis. Please add the same in else portion as well (when treatment is in progress, update the rate).

fixed.

fixed.

fixed

fixed

fixed

fixed

fixed

fixed

added

added

Could you update the Message ID to latest? 0XA9 is already used.

Could you update the Message ID to latest? 0XA9 is already used.

Align comment.

Align comment.

where are we calling an alarm? i think we are only sending event change data in completeBolusToCumulative() right?

where are we calling an alarm?
i think we are only sending event change data in completeBolusToCumulative() right?

do we need this block?

do we need this block?

update the comment as required: ml or L?

update the comment as required: ml or L?

This needs TD source update and also dialin update, keep informed system team about this change as this would impact DD demo running.

This needs TD source update and also dialin update, keep informed system team about this change as this would impact DD demo running.

Please add HDF Pump speed register and add a monitor in the substitution pump controller to compare against the target speed and if the difference is out of range, trigger an alarm.

Please add HDF Pump speed register and add a monitor in the substitution pump controller to compare against the target speed and if the difference is out of range, trigger an alarm.

Substitution pump

Substitution pump

Please add the min and max for tmp as well in the outputs.

Please add the min and max for tmp as well in the outputs.

rejReason = INVALID REQUEST FORMAT