leahi-fwcommon

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
LDT-3963 enabled fault mode tranisiton for switching off hetares for dilaysate flow rate testing

Do we have a return confirmation that the command is accepted by DD?

Do we have a return confirmation that the command is accepted by DD?

Is the UF rate unit in ml/min (instead of hr.)?

Is the UF rate unit in ml/min (instead of hr.)?

Are we assuming the sub mode return is pre-gen sub mode always?

Are we assuming the sub mode return is pre-gen sub mode always?

create the getDialGoodToDeliverStatus() function and have that return true rather than have the assignment here

create the getDialGoodToDeliverStatus() function and have that return true rather than have the assignment here

comment still needed?

comment still needed?

LEAHI-TD-FIRMWARE-LDT-4188_SW/FW integration to run a Treatment from UI
LEAHI-TD-FIRMWARE-LDT-4188_SW/FW integration to run a Treatment from UI
LDT-4188 updated pregen dialysate payload

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

LDT-473 merge latest into develop

LDT-473 merge latest into staging

LDT-473 merge latest from staging

???

???

Do we need a parameter for confirmation?

Do we need a parameter for confirmation?

First two logics ANDed correct (first cycle and close period). the third one (BC switch only must be ANDed only with close period), not to include first cycle..

First two logics ANDed correct (first cycle and close period). the third one (BC switch only must be ANDed only with close period), not to include first cycle..

LEAHI-DD-FIRMWARE-LDT-4000_HDF - FW Implementation - 1/4:
LEAHI-DD-FIRMWARE-LDT-4000_HDF - FW Implementation - 1/4:
It looks like you deleted it instead of fixing it.

It looks like you deleted it instead of fixing it.

Need to request syringe pump to start a seek before we go to seek state. It's not implemented yet, so just put a TODO comment for now.

Need to request syringe pump to start a seek before we go to seek state. It's not implemented yet, so just put a TODO comment for now.

Need to check heparin bolus volume > 0 too.

Need to check heparin bolus volume > 0 too.

Don't we need to check this too still?

Don't we need to check this too still?

Do we need to reset authStartTime here before we go to next state?

Do we need to reset authStartTime here before we go to next state?

Put positive case first ( TRUE == ...), then alarm and reject cases.

Put positive case first ( TRUE == ...), then alarm and reject cases.

No occlusion sensor in Leahi. Delete variable and references to it.

No occlusion sensor in Leahi. Delete variable and references to it.

We should now send the scanned text to the UI s/w for authentication before we go to next state where we wait for UI response.

We should now send the scanned text to the UI s/w for authentication before we go to next state where we wait for UI response.

Remove extra l at end of flag name (mis-spelled). And why are we sending this flag with scan event? Seems like the Received flag is what you want here.

Remove extra l at end of flag name (mis-spelled). And why are we sending this flag with scan event? Seems like the Received flag is what you want here.

The barcode reader is under TD f/w control via the FPGA (not implemented yet). So asking UI s/w to do a scan doesn't make sense.

The barcode reader is under TD f/w control via the FPGA (not implemented yet). So asking UI s/w to do a scan doesn't make sense.