LEAHI-DD-FIRMWARE-LDT-1477_Ultrafiltration - DD - 04: DEV - Feature Implementation
LEAHI-DD-FIRMWARE-LDT-1477_Ultrafiltration - DD - 04: DEV - Feature Implementation
LDT-1477 UF update

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 8 more files in changeset.
Done.

Done.

Updated.

Updated.

copy paste error comment

copy paste error comment

Please convert bool expression to BOOL (TRUE or FALSE) before assigning (either in the expression with trinary or prior).

Please convert bool expression to BOOL (TRUE or FALSE) before assigning (either in the expression with trinary or prior).

Done

Done

Done.

Done.

Add a comment to right listing math functions used.

Add a comment to right listing math functions used.

Remove blank line.

Remove blank line.

Looks crucible is not seeing those. let me check once with Dara.

Looks crucible is not seeing those. let me check once with Dara.

Done.

Done.

Updated.

Updated.

Updated

Updated

seems to be old , need to merge the staging code

seems to be old , need to merge the staging code

align the comment

align the comment

complicated check should be replaced with boolean flag

complicated check should be replaced with boolean flag

Variable name is not understood -temDataCollectionTimeInterval ?

Variable name is not understood -temDataCollectionTimeInterval ?

LDT-1473 latest from staging

  1. … 7 more files in changeset.
Yes. updated.

Yes. updated.

So shouldn't we stop trimmer heater here then?

So shouldn't we stop trimmer heater here then?

Done.

Done.

Done

Done

Since trimmer heater is enabled, no need of temp compensation. only for testing purposes, temp compensation is needed. Hence by default, temp compensation is disabled.

Since trimmer heater is enabled, no need of temp compensation. only for testing purposes, temp compensation is needed. Hence by default, temp compensation is disabled.

For spent chamber fill, we don't know how low the liquid level is ( in Chamber H) and hence decided not to turn on the trimmer heater for safety reasons. For bicarb fill, the dry bicart code should...

For spent chamber fill, we don't know how low the liquid level is ( in Chamber H) and hence decided not to turn on the trimmer heater for safety reasons. For bicarb fill, the dry bicart code should cover the actuators status. the current bicarb chamber fill code here should be updated with the newly implemented dry bi cart code changes.

I think cal data and count should be retrieved by same function and possibly needs some interrupt protection to ensure data and count stick together. If these functions are called from general task...

I think cal data and count should be retrieved by same function and possibly needs some interrupt protection to ensure data and count stick together. If these functions are called from general task, we could have situation where data is retrieved, then priority task updates fpga, and then counter is retrieved with wrong (incremented) counter.
If these two functions are being called from a driver that is executed in priority task, then we don't need to worry about this situation, but still ought to add a constraint/warning comment in function headers noting that these functions should only be called from priority task.