a time out for decaying D66 pressure to 3 PSI after D65 closed

a time out for decaying D66 pressure to 3 PSI after D65 closed

remove space after <= and add space end of function call.

remove space after <= and add space end of function call.

Is this time out after D80 opening or D65 closing state? if it is D65 closed, you need to start the timer after D65 closing.

Is this time out after D80 opening or D65 closing state? if it is D65 closed, you need to start the timer after D65 closing.

if we are not checking level, this can be removed and function name also can be updated accordingly.

if we are not checking level, this can be removed and function name also can be updated accordingly.

Is this a TODO? If so, add TODO to the comment and maybe be a little more descriptive of what needs to be done.

Is this a TODO? If so, add TODO to the comment and maybe be a little more descriptive of what needs to be done.

can we get 'runPIController' inside deadband control check?

can we get 'runPIController' inside deadband control check?

???

???

Do we need a parameter for confirmation?

Do we need a parameter for confirmation?

First two logics ANDed correct (first cycle and close period). the third one (BC switch only must be ANDed only with close period), not to include first cycle..

First two logics ANDed correct (first cycle and close period). the third one (BC switch only must be ANDed only with close period), not to include first cycle..

Can we check payload.index is equal to D6_LEVL (instead of above comparision)?

Can we check payload.index is equal to D6_LEVL (instead of above comparision)?

Remove space after 'Supply'.

Remove space after 'Supply'.

Align the value declaration indentation.

Align the value declaration indentation.

looks you are already checking this state in a line above.

looks you are already checking this state in a line above.

LDT-3978 fix override request for D6 and P25

Merge branch 'LDT-3978-switch-chamber-f-conductive-lev' of bitbucket.org:diality-cloud/leahi-common into LDT-3978-switch-chamber-f-conductive-lev

LEAHI-DD-FIRMWARE-LDT-4000_HDF - FW Implementation - 1/4:
LEAHI-DD-FIRMWARE-LDT-4000_HDF - FW Implementation - 1/4:
It looks like you deleted it instead of fixing it.

It looks like you deleted it instead of fixing it.

Need to request syringe pump to start a seek before we go to seek state. It's not implemented yet, so just put a TODO comment for now.

Need to request syringe pump to start a seek before we go to seek state. It's not implemented yet, so just put a TODO comment for now.

Need to check heparin bolus volume > 0 too.

Need to check heparin bolus volume > 0 too.

Don't we need to check this too still?

Don't we need to check this too still?

Do we need to reset authStartTime here before we go to next state?

Do we need to reset authStartTime here before we go to next state?

Put positive case first ( TRUE == ...), then alarm and reject cases.

Put positive case first ( TRUE == ...), then alarm and reject cases.

No occlusion sensor in Leahi. Delete variable and references to it.

No occlusion sensor in Leahi. Delete variable and references to it.

We should now send the scanned text to the UI s/w for authentication before we go to next state where we wait for UI response.

We should now send the scanned text to the UI s/w for authentication before we go to next state where we wait for UI response.