This needs TD source update and also dialin update, keep informed system team about this change as this would impact DD demo running.

This needs TD source update and also dialin update, keep informed system team about this change as this would impact DD demo running.

Please add HDF Pump speed register and add a monitor in the substitution pump controller to compare against the target speed and if the difference is out of range, trigger an alarm.

Please add HDF Pump speed register and add a monitor in the substitution pump controller to compare against the target speed and if the difference is out of range, trigger an alarm.

Substitution pump

Substitution pump

Missing comment.

Missing comment.

update comment. next one as well.

update comment. next one as well.

blank line after the banner and maintain alphabetical order of include files.

blank line after the banner and maintain alphabetical order of include files.

LDT-4001: previous CR comments. updates to TD interface and messaging

  1. … 3 more files in changeset.
LEAHI-DD-FIRMWARE-LDT-4001_HDF - FW Implementation - 2/4:
LEAHI-DD-FIRMWARE-LDT-4001_HDF - FW Implementation - 2/4:
LDT-4001: fixes and merges from previous working branch.

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

LDT-4188 merge latest from staging

LEAHI-DD-FIRMWARE-LDT-4246_FW Support for UI Demo
LEAHI-DD-FIRMWARE-LDT-4246_FW Support for UI Demo
LDT-4188 resolved compiler warning

LEAHI-DD-FIRMWARE-LDT-4188_SW/FW integration to run a Treatment from UI
LEAHI-DD-FIRMWARE-LDT-4188_SW/FW integration to run a Treatment from UI
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.
LDT-4188 addressed review comments

  1. … 1 more file in changeset.
To resolve the build failure with DD, haven't added all the features, instead added TODO comment

To resolve the build failure with DD, haven't added all the features, instead added TODO comment

Idea is to have this in institutional records we implement it

Idea is to have this in institutional records we implement it

If this is const, how will user add conversions (after 3rd value)?

If this is const, how will user add conversions (after 3rd value)?

Remove extra blank line.

Remove extra blank line.

Can we add sodium and bicarb overrides now, so that we don't want to come back again for this change? Also, please get the dialin updated to reflect the new parameters additions in tdproxy and let ...

Can we add sodium and bicarb overrides now, so that we don't want to come back again for this change?
Also, please get the dialin updated to reflect the new parameters additions in tdproxy and let system team be in loop to update the required system scripts?