AlarmDefs.h

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
LEAHI-TD-FIRMWARE-LDT-3215_Measured venous pressure decreases to atmosphere if H12 air pump is dead headed
LEAHI-TD-FIRMWARE-LDT-3215_Measured venous pressure decreases to atmosphere if H12 air pump is dead headed
Update alarms and Input/Output sections.

Update alarms and Input/Output sections.

Also, add option individual parameter override for substitutions pump rate as well.

Also, add option individual parameter override for substitutions pump rate as well.

Space after parenthesis. Please add the same in else portion as well (when treatment is in progress, update the rate).

Space after parenthesis. Please add the same in else portion as well (when treatment is in progress, update the rate).

fixed.

fixed.

fixed

fixed

fixed

fixed

fixed

fixed

added

added

where are we calling an alarm? i think we are only sending event change data in completeBolusToCumulative() right?

where are we calling an alarm?
i think we are only sending event change data in completeBolusToCumulative() right?

do we need this block?

do we need this block?

update the comment as required: ml or L?

update the comment as required: ml or L?

This needs TD source update and also dialin update, keep informed system team about this change as this would impact DD demo running.

This needs TD source update and also dialin update, keep informed system team about this change as this would impact DD demo running.

Please add HDF Pump speed register and add a monitor in the substitution pump controller to compare against the target speed and if the difference is out of range, trigger an alarm.

Please add HDF Pump speed register and add a monitor in the substitution pump controller to compare against the target speed and if the difference is out of range, trigger an alarm.

Substitution pump

Substitution pump

rejReason = INVALID REQUEST FORMAT

rejReason = INVALID REQUEST FORMAT

Missing comment.

Missing comment.

update comment. next one as well.

update comment. next one as well.

blank line after the banner and maintain alphabetical order of include files.

blank line after the banner and maintain alphabetical order of include files.

LEAHI-DD-FIRMWARE-LDT-4001_HDF - FW Implementation - 2/4:
LEAHI-DD-FIRMWARE-LDT-4001_HDF - FW Implementation - 2/4:
I think these remaining 4 else ifs need to have another && condition that your service is in idle state so that we don't change back into one of these 4 states while a bolus is in progress.

I think these remaining 4 else ifs need to have another && condition that your service is in idle state so that we don't change back into one of these 4 states while a bolus is in progress.

Can we restore the real code here yet? Ask Dara.

Can we restore the real code here yet? Ask Dara.

use ternary to make TRUE or FALSE.

use ternary to make TRUE or FALSE.

Fixed

Fixed

Yes

Yes

REmoved

REmoved