What if we're in paused state (because your alarm has stop property set)? Should we return TRUE here if in paused state and interrupted state was the await syringe load confirmation state?

What if we're in paused state (because your alarm has stop property set)? Should we return TRUE here if in paused state and interrupted state was the await syringe load confirmation state?

Right now, the load syringe alarm indicates stop so we will go to paused state. Is that what we want to do? Don't we want to stay in this state and wait for confirmation?

Right now, the load syringe alarm indicates stop so we will go to paused state. Is that what we want to do? Don't we want to stay in this state and wait for confirmation?

Should this alarm have stop property set to FALSE?

Should this alarm have stop property set to FALSE?

LEAHI-TD-FIRMWARE-LDT-4009_Isolated UF (TD) - FW Implementation - 1/2:
LEAHI-TD-FIRMWARE-LDT-4009_Isolated UF (TD) - FW Implementation - 1/2:
LEAHI-TD-FIRMWARE-LDT-4009_Isolated UF (TD) - FW Implementation - 1/2:
LEAHI-TD-FIRMWARE-LDT-4009_Isolated UF (TD) - FW Implementation - 1/2:
updated

updated

updated

updated

updated.

updated.

updated

updated

removed.

removed.

initialized using resetRinsebackFlags()

initialized using resetRinsebackFlags()

those are initialized in resetTreatmentRecircFlags() and that is called in init

those are initialized in resetTreatmentRecircFlags() and that is called in init

align comment

align comment

Can we please break this line as to match our standard of 100 columns? SET_ALARM_WITH_2_U32_DATA( ALARM_ID_TD_AIR_PUMP_STALL, currentAirPumpPowerLevel, getU32OverrideValue( &currentAirPumpRPM ) );

Can we please break this line as to match our standard of 100 columns?

SET_ALARM_WITH_2_U32_DATA( ALARM_ID_TD_AIR_PUMP_STALL,
currentAirPumpPowerLevel,
getU32OverrideValue( &currentAirPumpRPM ) );

use FALSE

use FALSE

use FALSE

use FALSE

It looks like the rows may be misaligned with the alarm ID declarations. For example, ALARM_ID_TD_BATTERY_CHARGER_ERROR_DETECTED is declared as 204.

It looks like the rows may be misaligned with the alarm ID declarations.
For example, ALARM_ID_TD_BATTERY_CHARGER_ERROR_DETECTED is declared as 204.

When we send any message to UI, usually add the message id in the doxygen header like this. @details \b Message \b Sent: MSG_ID_TD_RESP_TREATMENT_PARAMS_TO_VALIDATE Please refer to the function s...

When we send any message to UI, usually add the message id in the doxygen header like this.

@details \b Message \b Sent: MSG_ID_TD_RESP_TREATMENT_PARAMS_TO_VALIDATE

Please refer to the function sendTreatmentParamsResponse()

When we send any message to UI, usually add the message id in the doxygen header like this. *@details \b Message \b Sent: MSG_ID_TD_RESP_TREATMENT_PARAMS_TO_VALIDATE Please refer to the functi...

When we send any message to UI, usually add the message id in the doxygen header like this.

  • @details \b Message \b Sent: MSG_ID_TD_RESP_TREATMENT_PARAMS_TO_VALIDATE


Please refer to the function sendTreatmentParamsResponse()

When we send any message to UI, usually add the message id in the doxygen header like this. *@details \b Message \b Sent: MSG_ID_TD_RESP_TREATMENT_PARAMS_TO_VALIDATE Please refer to the functi...

When we send any message to UI, usually add the message id in the doxygen header like this.

  • @details \b Message \b Sent: MSG_ID_TD_RESP_TREATMENT_PARAMS_TO_VALIDATE


Please refer to the function sendTreatmentParamsResponse()

It looks like the rows may be misaligned with the alarm ID declarations. For example, ALARM_ID_TD_SAFETY_SHUTDOWN_POST_TEST_FAILED is declared as 207.

It looks like the rows may be misaligned with the alarm ID declarations.
For example, ALARM_ID_TD_SAFETY_SHUTDOWN_POST_TEST_FAILED is declared as 207.

Should Instructions part be this? A problem was detected with the hemodialysis device.

Should Instructions part be this?

A problem was detected with the hemodialysis device.

Please confirm if the source is DD?

Please confirm if the source is DD?